background image

User Manual

16

EverLight® 

(Post Mount)

EMC  Information

Table 2 

Guidance and manufacturer’s declaration – electromagnetic immunity 

The

EverLight

 is intended for use in the electromagnet environment specified below.  The 

customer or the end user of the 

EverLight

 should assure that it is used in such an environment. 

Immunity test 

IEC 60601 

Test level 

Compliance level 

Electromagnetic environment 

- guidance 

Electromagnetic 

discharge (ESD) 

IEC 61000-4-2 

+6 kV contact 

+8 kV air 

+6 kV contact 

+8 kV air 

Floors should be wood, 

concrete or ceramic tile. If 

floors are covered with 

synthetic material, the  

relative humidity should be at 

least 30 %. 

Electrical fast 

transient/burst 

IEC 61000-4-4 

+2 kV for power 

supply lines 

+1 kV for input/output 

lines

+2 kV for power 

supply lines 

+1 kV for input/output 

lines

Mains power quality should 

be that of a typical 

commercial or hospital 

environment. 

Surge

IEC 61000-4-5 

+1 kV differential 

mode

+2 kV common mode 

+1 kV differential 

mode

+2 kV common mode 

Mains power quality should 

be that of a typical 

commercial or hospital  

environment. 

Voltage dips, 

short 

interruptions and 

voltage

variations on 

power supply 

input lines 

IEC 61000-4-11 

<5 % U

T

(>95 % dip in U

T

)

for 0.5 cycle 

40 % U

T

(60 % dip in U

T

)

for 5 cycles 

70 % U

T

(30 % dip in U

T

)

for 25 cycles 

<5 % U

T

(>95 % dip in U

T

)

for 5 sec 

<5 % U

T

(>95 % dip in U

T

)

for 0.5 cycle 

40 % U

T

(60 % dip in U

T

)

for 5 cycles 

70 % U

T

(30 % dip in U

T

)

for 25 cycles 

<5 % U

T

(>95 % dip in U

T

)

for 5 sec 

Mains power quality should 

be that of a typical 

commercial or hospital 

environment.  If the user of 

the

EverLight

 requires 

continued operation during 

power mains interruptions, it 

is recommended that the 

Ever

Light

be powered 

from an uninterruptible

power supply or a battery. 

Power frequency 

(50/60 Hz) 

magnetic field 

IEC 61000-4-8 

3 A/m 

3 A/m 

Power frequency magnetic 

fields should be at levels 

characteristic of a typical 

location in a typical 

commercial or hospital 

environment. 

Summary of Contents for EverLight

Page 1: ...User Manual EverLight Post Mount ...

Page 2: ...e 5 6 Post Mount for Magellan CMU 6 Section III Operation Light Switch 7 Touch Pads 7 Positioning Light 8 Eliminating Adjustable Arm Drift 9 Section IV Care Cleaning Disinfecting 10 SectionV User Service Information Troubleshooting 11 Service Instruction 11 Disposal of Equipment 11 SectionVI Parts List Diagrams Light Arm 12 Light Head 12 Limited Warranty 13 Wiring Schematic 14 EMC Information 15 1...

Page 3: ... kg 27 lbs The authorized European representative is DentalEZ GB Ltd Cleveland Way Hemel Hempstead Hertfordshire HP2 7DY England Phone 01442 269301 Attn Mr Andy King This manual contains installation operation and care instructions and user service information for the EverLight LED operatory light The EverLight operatory light system is intended to be used by trained professional dental care perso...

Page 4: ... be installed according to EMC information See EMC Information Pgs 15 through 18 Mobile radio frequency RF communications equipment can effect medical electrical equipment Section I Introduction Caution Warning Biohazard Warning DangerousVoltage Alternating Current Direct Current Type B Applied Part Protective Earth ground General Mandatory Action Refer to Manual follow instructions European Certi...

Page 5: ...omponents are ordered Carefully unpack the light Inspect all of the items received to assure there has been no shipping damage If damage is discovered contact the carrier immediately Then report and file the appropri ate claims with the carrier within 24 hours after delivery See Limited Warranty Page 13 Light Arm Assembly Light Post Post Mount Arm Assembly Dimensions 15 17 7 25 2 24 31 75 1 701 Pa...

Page 6: ... the light arm assembly into the bushing until it is fully seated in the light post 8 Using a 5 64 Allen wrench turn the set screw in the light post until it passes through the light arm assembly spindle slot and engages the light arm assembly Then loosen the set screw two revolutions for clearance Power Supply Box 1 Inside the post mounted service console plug the operatory light cord into the po...

Page 7: ...x 1 Plug the supplied power cord into the power supply See photo below 2 Plug the light harness from the light arm into the power supply 3 Route the supplied power cord harness through the front of the seat casting and plug it into the chair harness See second photo on this page 4 Roll up any excess wire and neatly place them inside the seat casting making sure no wires will interfere with moving ...

Page 8: ...Composite Setting Sun 5800 Kelvin Color Matching Procedural Settings Intensity Settings Intensity Settings 1 Low 15 000 Lux 2 Medium 25 000 Lux 3 High 30 000 Lux Medium 25 000 Lux High 30 000 Lux The light ON OFF toggle switch is located on the elbow section of the light arm assembly Toggle Switch Flip the toggle switch to the UP to turn the light ON Flip the toggle switch DOWN to shut the light O...

Page 9: ...ither side of the LED head position the light to illuminate the oral cavity See Light Head Range Of Motion Pg 3 Handles Positioning Light To take full advantage of the EverLight s intensity and color correction position the light between 20 and 45 away from the oral cavity NOTICE ...

Page 10: ...ong with the end cap 3 Insert a 9 64 Allen wrench into the two small holes in the side of the adjustable arm and completely loosen the friction block 4 Insert a large flat head screwdriver into the large slot in the side of the adjustable arm and tighten the slotted nut 5 Adjust the friction block to provide one to two pounds maximum of break away force 6 Replace the adjustable arm cover and secur...

Page 11: ...soap and water or a neu tralizer for iodophors may reduce discoloration Cleaning Disinfecting Under normal use there is no routine maintenance required NOTICE Avoid spraying anything directly on the light Spray cloth first then wipe CAUTION Do not use alcohol based disinfectants on plastic surfaces CAUTION All surface disinfectants cause varying degrees of discoloration to light colored surfaces C...

Page 12: ...mber Plate The chart below should be used when troubleshoot ing EverLight problems If these suggested trouble shooting procedures do not resolve the problem see the service instruction that follows Disposal and Decommissioning of DentalEZ product Note All local regulatory requirements for disposal and decommissioning of equipment apply Electrical Salvage Remove all circuit board and electrical cab...

Page 13: ...29 7 Left Side LED Light Control 3802 328 8 Color Control Decal 3802 337 9 Intensity Control Decal 3802 336 10 Light Control Housing Kit w Decals 3802 338 11 Front Clear Cover Set 3802 331 Ribbon Cable Kit 3802 332 Power Supply Kit 3802 334 Power Supply Kit CE 3802 352 Complete Head 3802 354 Light Head Part Kit Name Part Kit No 1 Arm End Cap 3801 811 2 Rigid Arm Cap 3801 814 3 Post Bushing 3801 84...

Page 14: ...s Light Components Warranty Period Structural Members Fixed Arm Adjustable Arm and Handles 3 Years from date of installation Switches LED Driver Lens LED 6 Years from date of installation Please note the following additional terms of our warranty and return policy Warranties cover manufacturing defects only and do not cover defects resulting from abuse improper handling cleaning care or maintenanc...

Page 15: ...User Manual 14 EverLight Post Mount Wiring Schematic Power Supply INPUT 100 240VAC 50 60 HZ OUTPUT 12VDC 40 W White Neutral Green Ground Black Line ...

Page 16: ...not likely to cause any interference in nearby electronic equipment RF Emissions CISPR 11 Class A The EverLight is suitable for use in all establishments other than domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes Harmonic emissions IEC 61000 3 2 Class A Not Applicable Voltage fluctuations flicker...

Page 17: ...ial or hospital environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 ...

Page 18: ...of transmitter m Rated maximum output power of transmitter W 150 kHz to 80 MHz P v d 1 5 3 80 MHz to 800 MHz P E d 1 5 3 800 MHz to 2 5 GHz P E d 1 7 0 01 0 12 0 12 0 23 0 1 0 34 0 34 0 74 1 1 7 1 7 2 3 10 3 7 3 7 7 4 100 11 7 11 7 23 3 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable ...

Page 19: ... from fixed RF transmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption ...

Page 20: ... DentalEZ Alabama Inc Printed in USA PN 2717 252B November 2015 2500 Highway 31 South Bay Minette Alabama 36507 866 DTE INFO Fax 251 937 0461 www dentalez com ...

Reviews: