11
Section V User Service Information
No light
Master/light switch is in
OFF
position Turn
ON
master/light switch.
No power
Check circuit breaker.
Power supply failure
Replace power supply.
Broken arm wire harness
Check arm wire harness for
continuity.
Faulty switch
Replace switch.
Low light intensity
Intensity set on low
Set intensity to high.
Plastic light shield dirty
Clean light shield.
Light arm drifts vertically
Counterbalance out of adjustment
Adjust arm drift.
(See Page 9.)
Light arm drifts horizontally
Light post not level
Level light post.
If the area of concern is not addressed in this man-
ual, contact your local DentalEZ full-service dealer-
ship.
(See Limited Warranty, Page 13.)
Please have the following product information
available:
•
Model Name ___________________________
•
Model Number _________________________
•
Serial Number __________________________
•
Installation Date ________________________
•
Dealer ________________________________
Symptom
Possible Cause(s)
Solution
Light Arm
Assembly
Model/Serial
Number Plate
The chart below should be used when troubleshoot-
ing EverLight problems. If these suggested trouble
-
shooting procedures do not resolve the problem, see
the service instruction that follows.
Disposal and Decommissioning of DentalEZ product
Note: All local regulatory requirements for
disposal and decommissioning of equipment apply.
Electrical Salvage: Remove all circuit board and
electrical cabling for recycle as electrical salvage.
Metal Salvage: Remove all aluminum and steel
components for recycle as metal salvage.
Plastic Salvage: Remove all plastic components for
recycle as plastic salvage.
Non-Salvage Components: All other material
unsuitable for recycling should be disposed of
properly.
Troubleshooting
Service Instruction
Disposal of Equipment
Exercise extreme caution when
troubleshooting the electrical components
of the EverLight. When electrical power
is required, safety precautions
must
be
observed.
WARNING
For convenient referencing, a color-coded wiring
schematic is located at the end of this manual.
NOTICE
The DentalEZ Technical Service Department is
available to supply service personnel with any
additional information or instructions needed to
repair or maintain the dental light.
NOTICE