HD2402
- 43 -
V1.4
6.4.8
Table m.n.o.
380-3000
Table
m.n.o. 380-3000
shows the exposure limits for parameters “
m
”, “
n
”, “
o
”.
The displayed values are:
•
E
skin
[W/m
2
] irradiance value (corresponding to the instant value if a single
sample is taken into consideration, and to the mean value if several samples
are taken into consideration).
•
H
skin
[J/m
2
] dose value regarding the samples taken into consideration.
•
Exposure time limit (t_lim) for “
o
” index necessary for exceeding the threshold,
calculated with the following formula:
t_lim = 35600/E
skin
if the found time is > 10s
otherwise:
t_lim = (20000/ E
skin
)
4/3
•
Exposure time limit (t_lim) for indexes “
m
” and “
n
”
Limits “
m
” and “
n
” should be calculated in the 780-3000 spectral interval. Since this
spectral range is not available in the instrument, the same channel of the “
o
” limit is
used; in this case, values will be greater than real values and consequently they will
be in favor of worker safety.
As to exposure times greater than 1000 seconds, the “
n
” limit is calculated:
if E
IR
≤
100 W/m
2
then
the
“
n
” limit is not satisfied;
if E
IR
> 100 W/m
2
then the exposure time is calculated in order to return back with-
in “
n
” limit, in particular:
t_lim=(18000/
E
IR
)
4/3
If the
Include in Report
box is selected, the data in the table will be included in the fi-
nal report.