![DeltaOHM HD2402 Operating Manual Download Page 27](http://html1.mh-extra.com/html/deltaohm/hd2402/hd2402_operating-manual_2479058027.webp)
HD2402
- 27 -
V1.4
6.4.3
Table a. 180-400
Table
a. 180-400
describes the exposure limit for parameter “
a
”.
The displayed data are:
•
Effective irradiance value E
eff
[W/m
2
] (corresponds to the instant value if a sin-
gle sample is considered, or to the mean value if several samples are consid-
ered).
•
Effective dose value H
eff
[J/m
2
] regarding the considered samples.
•
Time in hours/minutes/seconds for exceeding 30 J
eff
/m
2
maximum daily dose
calculated starting from the irradiance mean value.
If the
Include in Report
box is selected, the data in the table will be included in the fi-
nal report.
6.4.4
Table b. 315-400
Table
b. 315-400
shows the exposure limit for parameter “
b
”.
The displayed data are:
•
Irradiance value E
UVA
[W/m
2
] (corresponds to the instant value if a single sam-
ple is considered, or to the mean value if several samples are considered).
•
Dose value H
UVA
[J/m
2
] regarding the considered samples.
•
Time in hours/minutes/seconds for exceeding 10000 J
UVA
/m
2
maximum daily
dose calculated starting from the irradiance mean value.
If the
Include in Report
box is selected, the data in the table will be included in the fi-
nal report.