HD2402
- 18 -
V1.4
6
MANAGEMENT OF A MEASUREMENT CAMPAIGN WITH
DELTALOG13
The chapter describes all steps needed to perform a measurement campaign, with the
purpose to evaluate and analyze one or more sources and positions in the same place.
A measurement campaign is performed through the following steps:
1.
Setting measurement scales for the different channels;
2.
Setting mode and duration of the measurement log;
3.
Measurement log;
4.
Download to a PC of the acquired measures, if these were stored in the in-
ternal memory of the instrument.
5.
Analysis of data and issue of evaluation report.
6.1
S
ETTING MEASUREMENT SCALES
Before every reading it is of the utmost importance to check that the proper meas-
urement ranges are set in the instrument. If the set measurement range is too small,
measurements can be beyond full scale, thus not allowing a proper evaluation of the
readings.
4 measurement scales are available for all the channels, except for channel 9, that
has only one measurement scale (Table 3 at page 11).
In order to set measurement ranges, connect the instrument to a PC, start Delta-
Log13 and wait for the automatic identification of the instrument. Select the
Instru-
ment Setup
icon and then the
Input setting
item. The following window will appear:
The following values are displayed for each measurement channel, except for channel
9: measured value (mean value per second), peak value (per second), preset meas-
urement scale and current scale, that is the scale in use at the moment of the last ac-
quisition.
The preset scale is in light blue color. All scales are factory preset to
AUTO
(auto-
matic). Scales for each channel can be set manually, by mouse-clicking on the desired
scale (the smallest bar indicates the full sensitivity scale).