HD2402
- 37 -
V1.4
if Alpha > 100mrad it is assumed that C
α
= 100
if L
R_real
≤
2.8
•
10
5
⇒
t > 10s (g limit satisfied)
if 2.8
•
10
5
≤
L
R_real
≤
8.89
•
10
6
⇒
t_lim
=
(5
•
10
5
/L
R_100
)
4
[s]
(h limit satisfied)
if 8.89
•
10
6
≤
L
R_real
⇒
t_lim < 10
μ
s (i limit)
if 1.7mrad
≤
Alpha < 100mrad it is assumed that C
α
= Alpha
if L
R_real
≤
2.8
•
10
7
/Alpha
⇒
t > 10s (g limit satisfied)
if 2.8
•
10
7
/Alpha
≤
L
R_real
≤
8.89
•
10
8
/Alpha
⇒
t_lim
=
(5
•
10
7
/(L
R_real
•
Alpha))
4
[s]
(h limit satisfied)
if 8.89
•
10
8
/Alpha
≤
L
R_real
⇒
t_lim < 10
μ
s (i limit )
•
Diameter d [m] – Distance R [m]
:
o
Alpha [rad] angle subtended by the source.
o
Omega [sr] solid angle subtended by the source.
o
L
R
(100mrad) [W/(m² sr)] radiance calculated assuming that the source angular
dimension is 100mrad.
o
L
R
(1.7mrad) [W/(m² sr)] radiance calculated assuming that the source angular
dimension is 1.7mrad.
o
L
R
(real) [W/(m² sr)] real radiance calculated starting from the entered geomet-
ric parameters.
o
t_lim [s] exposure time limit, that is the time period during which a source can
be observed without damages to the visual system. This value is calculated with
the following procedure:
if Alpha > 100mrad it is assumed that C
α
= 100
if L
R_real
≤
2.8
•
10
5
⇒
t > 10s (g limit satisfied)
if 2.8
•
10
5
≤
L
R_real
≤
8.89
•
10
6
⇒
t_lim
=
(5
•
10
5
/L
R_100
)
4
[s]
(h limit satisfied)
if 8.89
•
10
6
≤
L
R_real
⇒
t_lim < 10
μ
s (i limite)
if 1.7mrad < Alpha < 100mrad it is assumed that C
α
= Alpha
if L
R_real
≤
2.8
•
10
7
/Alpha
⇒
t > 10s (g limit satisfied)
if 2.8
•
10
7
/Alpha
≤
L
R_real
≤
8.89
•
10
8
/Alpha
⇒
t_lim
=
(5
•
10
7
/(L
R_real
•
Alpha))
4
[s]
(h limit satisfied)
if 8.89
•
10
8
/Alpha
≤
L
R_real
⇒
t_lim < 10
μ
s (i limit)
if Alpha < 1.7mrad it is assumed that C
α
= 1.7
if L
R_1.7
≤
1.65
•
10
7
⇒
t > 10s (g limit satisfied)
if 1.65
•
10
7
≤
L
R_1.7
≤
5.23
•
10
8
⇒
t_lim = (2.94
•
10
7
/(L
R_1.7
))
4
[s] (h limit satis-
fied)
if 5.23
•
10
8
≤
L
R_1.7
⇒
t_lim < 10
μ
s (i limit)
The exposure limit time in seconds is displayed aside radiance values.
If the
Include in Report
box is selected, table data will be included in the final report.