23
SERIAL
INTERFACE
AND
USB
The
HD21...4.2
instruments are fitted with an electrically isolated RS-232C serial interface, and an
USB 2.0 interface.
The following serial cables can be used:
HD2110CSNM
:
serial connection cable with 8-pole MiniDin connector on one end and
9-pole Sub D connector on the other end;
C.206
: serial connection cable with 8-pole MiniDin connector on one end and USB type A
connector on the other end. With integrated RS232/USB converter;
CP23
:
connection cable with Mini-USB type B connector on one end and USB type A
connector on the other end.
The connection via the C.206 cable requires the previous installation of the cable USB drivers.
Install the drivers
before connecting the C.206 cable to the PC
.
The connection via the CP23 cable does not require the installation of USB drivers: when connecting
the instrument to the PC, the Windows® operating system automatically recognizes the device as an
HID device (Human Interface Device) and uses the drivers already included in the operating system.
Cable
Instrument
port
PC
port
Installation of
USB drivers
HD2110CSNM
RS232 (MiniDin)
RS232 (9-pole SubD)
No
C.206
RS232 (MiniDin)
USB
Yes
CP23
USB (Mini-USB)
USB
No
The standard serial transmission parameters of the instrument are:
Baud rate 38400 baud
Parity
None
N. bit
8
Stop bit
1
Protocol
Xon/Xoff
It is possible to change the RS232C serial port baud rate by setting the "
Baudrate
" parameter in the
menu (please see page 11). The possible values are: 38400, 19200, 9600, 4800, 2400, 1200. The
other transmission parameters are fixed.
The USB 2.0 connection does not require the setting of parameters.
The instruments are provided with a complete set of commands and data queries to be sent via the PC.
The serial commands work with a standard serial communication program (e.g. Hyperterminal) only
through the RS232 serial port of the instrument, using the cable HD2110CSNM or the cable C.206.
All the commands transferred to the instrument must have the following structure:
XYcr
where:
XY
is the command code and
cr
is the Carriage Return (ASCII 0D)
Command
Response
Description
P0
&
Ping (locks the instrument keyboard for 70 seconds)
P1
&
Unlocks the instrument keyboard
S0
101.5E+3 22.7
Captured measurements (24 characters)
G0
Model HD2114 -2
Instrument model