26
The same is true for the Max-Min-Avg function.
If the logging is started when the display is in REL mode, the relative values are logged.
If the logging is started when the display is in Peak mode, the peak values are logged, provided
that the logging interval is equal to 1 second. Please see the paragraph dedicated to that on page
12.
It is possible to activate both the logging (LOG) and direct transfer (PRINT) functions at the
same time.
T
HE
FUNCTION
-
ONLY FOR
HD21...4.2
The PRINT function sends the measurements taken in real time by the instrument inputs directly to
a PC or a printer. Print data units of measurements are the same as those used on the display. The
function is started by pressing SERIAL. The time interval between two consecutive prints can be set
from 1 second to 1 hour (please see the
Print and log interval
menu item on page 10). If the print
interval is equal to 0, by pressing SERIAL the single data is sent to the connected device. If the
print interval is higher than 0, the data transfer continues until the operator stops it by pressing
SERIAL again.
The PRINT function works with a standard serial communication program (e.g. Hyperterminal) only
through the RS232 serial port of the instrument, using the cable HD2110CSNM or the cable C.206.
Connect the
HD40.1
printer using cable HD2110CSNM.
NOTES:
The print out is formatted across 24 columns.
Some keys are disabled during serial transmission. The following keys work: ON/OFF, HOLD,
FUNC (Max-Min-Avg) and LOG.
Pressing the HOLD, REL FUNC and PEAK keys has no effect on the printed data if these keys
are pressed
after
starting the printing, otherwise the following is valid.
The serial transfer started with the display in HOLD mode proceeds normally with the actual
measured values (that is, not in “HOLD” mode). Only the display is frozen to the values present
when the HOLD key was pressed.
The same is true for the Max-Min-Avg function.
If the serial transfer is started when the display is in REL mode, the relative values are
transferred.
If the serial transfer is started when the display is in Peak mode, the peak values are transferred,
provided that the printing interval is equal to 1 second. Please see the paragraph dedicated to
that on page 12.
It is possible to activate both the logging (LOG) and direct transfer (PRINT) functions at the
same time.