PP471

 MODULE FOR PRESSURE MEASUREMENT

 

The PP471 module acts as an interface between the TP704 and TP705 series Delta Ohm probes and 
the instrument.  
The TP704 series pressure probes have a 1/4” BSP threaded male connection and must be screwed 
in  the joint  where  you wish  to measure the pressure. The TP  705 series  have two Ø5 connections 
where  the  suitable  tubes  are  inserted  in  order  to  perform  the  desired  measurement.  Please  pay 
careful attention to the joint pressure sealing; use suitable gaskets and joints. A plastic cap protects 
the  threaded  connection.  Put  it  back  after  use  as  it  protects  the  pressure  cell  from  foreign  bodies. 

VERY IMPORTANT: ensure the probe's bottom scale is higher than the pressure that will be 
measured. In case this value is unknown, start by using higher-capacity probes. 

For  every  pressure  probe  a  range  of  overload  pressure  and  a  burst  pressure  are  declared: 
pressures  in  the  range  of  overload  pressure  don’t  cause  the  break  of  the  sensor  but  the 
declared  accuracy  can  be  overcome.  Pressure  over  the  stated  burst  pressure  may  break  the 
sensor. The application of higher pressures than the overload pressure limit, although lower 
than the burst pressure, may produce permanent damage to the probe (e.g. offset shift). Never 
exceed the stated burst pressure. 

Please see the technical characteristics of the probes on page 31.  

Upon  turning  on  the  instrument  automatically  detects  the  PP471  module.  The  probe's  type 
(absolute, relative or differential) and bottom scale value are detected even when the instrument is 
on.  If  no  logging  or  recording  operations  are  in  progress,  it  is  possible  to  change  the  probe 
connected to the module without turning the instrument off and on again.  

The measurement provided can be the instantaneous  or the peak value of the pressure: please see 
the PEAK key description on page 12. 
 

When the PP471 pressure module is connected, the internal sensor is disabled. 

 

D

IRECT INPUT INTO 

Pt100,

 

Pt1000

 AND 

Ni1000

 

T

EMPERATURE PROBES

 

The  instrument  accepts  the  input  of  Platinum  temperature  probes  with  resistances  of  100Ω  and 
1000Ω, as well as Nickel temperature probes with a 1000Ω resistance. 
The Pt100 are connected to 4 wires, the Pt1000 and Ni1000 to 2 wires, with the excitation current 
chosen minimizing the sensor self-heating effects. 
All  probes  with  module are  calibrated  in  the  factory,  the  2  or  4-wire  probes  with  direct  input 

are 

checked for conformity with class A tolerance 

according to norm IEC751 - BS1904 - DIN43760.  

For  the  temperature  probes  without  SICRAM  module  (Pt100  4  wires,  Pt1000  and  Ni1000)  the 
model configuration is required (please see the description of the Probe Type menu on page 10). 
 
The °C or °F unit of measurement can be chosen for display, printing, and logging using the °C/°F-
ESC key. 
 

How to measure 

The  response  time  for  the  measurement  of  the  temperature  in 

air 

is  greatly  reduced  if  the  air  is 

moving.  If  the  air  is  still,  stir  the  probe.  Please  remember  that  the  response  times  in  any  case  are 
longer than those for liquid measurements. 
 
The temperature measurement by 

immersion

 is carried out by inserting the probe in the liquid for at 

least 60mm; the sensor is housed in the end part of the probe. 

Summary of Contents for HD2114.0

Page 1: ...the instrument that you have purchased We cannot entirely exclude errors in the manual for which we apologize The data figures and descriptions contained in this manual cannot be legally asserted We reserve the right to make changes and correction without prior notice HD2114 0 HD2114 2 HD2134 0 HD2134 2 HD2164 0 HD2164 2 HD2114B 0 HD2114B 2 ENGLISH REV 2 2 02 07 2014 ...

Page 2: ...ACE AND USB 23 STORING AND TRANSFERRING DATA TO A PERSONAL COMPUTER 25 THE LOGGING FUNCTION ONLY FOR HD21 4 2 25 CLEARING THE MEMORY ONLY FOR HD21 4 2 25 THE PRINT FUNCTION ONLY FOR HD21 4 2 26 CONNECTION TO A PC 27 CONNECTION TO THE RS232C SERIAL PORT OF THE INSTRUMENT ONLY FOR HD21 4 2 27 CONNECTION TO THE USB 2 0 PORT OF THE INSTRUMENT ONLY FOR HD21 4 2 27 NOTES ABOUT WORKING AND OPERATIVE SAFE...

Page 3: ... instruments are dataloggers They memorize up to 36 000 samples which can be transferred from the instrument connected to a PC via the RS232C serial port or USB 2 0 port The storing interval printing and baud rate can be configured using the menu The HD2114 2 HD2134 2 HD2164 2 and HD2114B 2 instruments can transfer via the RS232C serial port the acquired measurements to a PC or to a portable print...

Page 4: ... 4 HD2114 0 HD2134 0 HD2164 0 Micromanometer Thermometer HD2114B 0 Barometer Thermometer ...

Page 5: ...ifference between the current value and the logged value when the key is pressed in the menu decreases the current value 9 PEAK key starts and ends the peak measurement 10 MENU key allows access to and exit from the menu 11 C F ESC key changes the unit of measurement for temperature from degrees Celsius to Fahrenheit in the menu cancels the operation in progress without making changes 12 UNIT key ...

Page 6: ... 6 HD2114 2 HD2134 2 HD2164 2 Micromanometer Thermometer HD2114B 2 Barometer Thermometer ...

Page 7: ... DUMP LOG key during normal operation starts and ends the saving of the data in the internal memory in the menu starts the data transfer from the instrument s memory to the PC 12 Mini USB type B connector for USB 2 0 For the connection to PC with cable CP23 13 PEAK MENU key during normal operation starts and ends the peak measurement When pressed together with the FUNC ENTER key allows access to t...

Page 8: ...ent therefore apply to the internal sensor unless the external PP471 module is connected Please see the paragraph dedicated to calibration on page 14 Each key specific function is described in detail below ON OFF Auto OFF key The instrument is turned on and off using the ON OFF key The turning on enables all display segments for a few seconds starts an auto test including the detection of the poss...

Page 9: ...irms the current parameter and then goes to the next one Pressed together with the PEAK MENU key opens the menu please see the PEAK MENU key description Attention the data captured using the Record function cannot be transferred to the PC HOLD key It increases the current parameter when used in the menu when used in measurement mode it freezes the measurement in progress and upon application of pr...

Page 10: ... menu items are listed in this order 1 Differential probes resetting In the differential probes there can be a slight difference between the two inputs so the instrument does not indicate zero at comparable pressure applied to the two inputs of a probe This is the reason the reset command was designed for the differential value leave the inputs of each probe open so that they detect the same press...

Page 11: ...r is started when the SERIAL key is pressed To end the recording LOG and continuous data transfer operations SERIAL with an interval greater than 0 press the same key again 6 Sleep_Mode_LOG Automatic turning off during recording only HD21 4 2 this function controls the instrument s automatic turning off during logging occurring between the capture of a sample and the next one When the interval is ...

Page 12: ...f previous measurements using FUNC CLR please see the FUNC ENTER key description IMPORTANT the peak measurement works with external probes connected to the instrument through a SICRAM PP471 module and not with the internal pressure sensor The Logging function during peak measurement only HD21 4 2 To record the maximum peaks start the peak function and then the record function LOG key As the measur...

Page 13: ... the blinking of the battery symbol when using an external power supply the battery symbol does not appear Press SERIAL to end the continuous printing Before starting the printing via the RS232C port set the baud rate To do so select the Baud Rate menu item and select the maximum value equal to 38400 baud by using the arrows Confirm by pressing ENTER The DeltaLog9 software for PC will automaticall...

Page 14: ...nt is enabled using these probes To change the instantaneous or peak value unit of measurement press UNIT The following units of measurement are available Pa hPa kPa mbar bar atm mmHg mmH2O kgf cm2 PSI inchHg Some units of measurement require a multiplication factor the 3 symbol indicates the displayed value must be divided by 1 000 the 3 and 6 symbols indicate the displayed value must be multipli...

Page 15: ...tom scale value are detected even when the instrument is on If no logging or recording operations are in progress it is possible to change the probe connected to the module without turning the instrument off and on again The measurement provided can be the instantaneous or the peak value of the pressure please see the PEAK key description on page 12 When the PP471 pressure module is connected the ...

Page 16: ...4 2 instruments also work with direct 4 wire Pt100 Pt1000 and Ni1000 probes manufactured by other producers for the instrument connection is prescribed the TP47 connector to which the probe s wires should be welded The instructions to connect the Platinum or Nickel probe to the module are provided below The module is supplied complete with fairlead and gasket for 5mm maximum diameter cables Do the...

Page 17: ...sure the cable is not twisted while you are screwing the fairlead Now the probe is ready Direct connection of 4 wire Pt100 sensors 4 wire Pt100 sensors can be soldered directly to the pins of the free female connector without making use of the TP47 board The 4 wires of the Pt100 sensors have to be soldered as indicated in the figure on the left In order to use this type of probe it is necessary to...

Page 18: ...nty and depends on the ability of the operator 9 Temperature measurements on non metal surfaces usually require a great deal of time due to the low heat conductivity of non metal materials 10 Probes are not insulated from their external casing be very careful not to come into contact with live parts above 48V This could be extremely dangerous for the instrument as well as for the operator who coul...

Page 19: ...ange For the pressure the limit is set to 120 of the bottom scale nominal value LOG MEM FULL Memory full the instrument cannot store further data the memory space is exhausted NEW PROBE DET This message appears when a probe is inserted into a functioning instrument Turn the instrument off and then back on again PROB ERR A probe with SICRAM module has been inserted when not admissible for that spec...

Page 20: ...ge values clearing FUNC CLRD max min and average values cleared HOUR hour LOG In PROG logging in progress LOG MEM FULL memory full LOG_CLRD memory data cleared LOG_STOP logging complete LOG_STRT logging started MIN USE_UNIT_TO_ZERO SEC minutes use the UNIT key to reset the seconds MNTH month NEW_PROB_DET new probe detected OVER maximum limit of temperature or pressure exceeded PLS_EXIT FUNC RES_FO...

Page 21: ... batteries in order to turn the instrument back on In order to avoid data loss the logging session is ended if the HD21 4 2 are logging and battery voltage falls below the minimum operating level In the HD21 4 2 the battery symbol turns off when the external power supply is connected To replace the batteries switch the instrument off and unscrew the battery cover counter clockwise After replacing ...

Page 22: ...ing Always use good quality leakproof alkaline batteries Sometimes on the market it is possible to find new batteries with an insufficient charge capacity INSTRUMENT STORAGE Instrument storage conditions Temperature 25 65 C Humidity less than 90 RH without condensation Do not store the instrument in places where Humidity is high The instrument may be exposed to direct sunlight The instrument may b...

Page 23: ...strument port PC port Installation of USB drivers HD2110CSNM RS232 MiniDin RS232 9 pole SubD No C 206 RS232 MiniDin USB Yes CP23 USB Mini USB USB No The standard serial transmission parameters of the instrument are Baud rate 38400 baud Parity None N bit 8 Stop bit 1 Protocol Xon Xoff It is possible to change the RS232C serial port baud rate by setting the Baudrate parameter in the menu please see ...

Page 24: ...uto power off function ENABLE KQ Auto power off function DISABLE KZ Differential probes reset function RA Reading of LOG PRINT interval set RP 600 Battery level Resolut 0 01V RUA U Pa Main unit of measurement RUB U C Secondary unit of measurement WA Setting LOG PRINT interval is a hexadecimal number 0 D that represents the position of the interval in the list 0 1 5 10 3600 seconds WC0 Setting SELF...

Page 25: ...g message LOG ON The data stored in the memory can be transferred to a PC using the DUMP LOG command press simultaneously the FUNC ENTER and PEAK MENU keys using ENTER select the _LOG_DUMP_or_ERAS item and then press on the LOG DumpLog key During data transfer the display shows the message DUMP to stop the data transfer press ESC on the instrument or on the PC CLEARING THE MEMORY ONLY FOR HD21 4 2...

Page 26: ...pressing SERIAL again The PRINT function works with a standard serial communication program e g Hyperterminal only through the RS232 serial port of the instrument using the cable HD2110CSNM or the cable C 206 Connect the HD40 1 printer using cable HD2110CSNM NOTES The print out is formatted across 24 columns Some keys are disabled during serial transmission The following keys work ON OFF HOLD FUNC...

Page 27: ...r USB serial port of the PC 3 Turn on the instrument and set the baud rate to 38400 MENU ENTER until the Baud Rate parameter select 38400 using the arrows confirm with ENTER The parameter remains in the memory until replacement of the batteries 4 Launch the DeltaLog9 application and press CONNECT Wait for the connection to occur and follow the indications on the screen For a description of the Del...

Page 28: ... safety of the instrument can only be guaranteed under the environmental and electrical operating conditions that are in specified in chapter TECHNICAL CHARACTERISTICS Do not use or store the product in places such as listed below Rapid changes in ambient temperature which may cause condensation Corrosive or inflammable gases Direct vibration or shock to the instrument Excessive induction noise st...

Page 29: ...240Vac 12Vdc 1A Measuring units C F Pa hPa mbar bar atm mmHg mmH2O kgf cm2 PSI inchHg Security of memorized data Unlimited independent of battery charge conditions Time Date and time Schedule in real time Accuracy 1min month max drift Measured values storage models HD21 4 2 Type 2000 pages containing 18 samples each Quantity 36000 samples pressure temperature Selectable storage interval 1s 5s 10s ...

Page 30: ...cale 20mbar 200mbar 2000mbar 600 1100mbar Maximum overpressure 300mbar 1bar 6bar 3bar Resolution 0 001mbar 0 01mbar 0 1mbar 0 1mbar Accuracy 23 C 0 3 f s 0 1 f s 0 1 measurement 0 3mbar Operating temperature 0 60 C Connection quick couplings Ø5mm Compensation temperature 0 60 C Drift on zero 1 f s 0 5 f s 0 5 f s 0 3 f s Drift on span 1 f s 0 5 f s 0 5 f s 0 3 f s Fluid contacting the membrane non...

Page 31: ... Tube 5mm 400 mbar 450 mbar TP704 200MBGI 0 25 FSO 10 80 C BSP 400 mbar 800 mbar 900 mbar 0 1 mbar TP704 400MBGI 0 25 FSO 10 80 C BSP 500 mbar 1500 mbar 1800 mbar 0 1 mbar TP705 500MBD 0 25 FSO 0 60 C Tube 5mm 600 mbar 1200 mbar 1500 mbar 0 1 mbar TP704 600MBGI 0 25 FSO 40 125 C BSP 1 00 bar 3 bar 3 3 bar 1 mbar TP705 1BD 0 25 FSO 0 60 C Tube 5mm 2 bar 2 2 bar TP705BARO 0 25 FSO 0 60 C Tube 5mm 2 ...

Page 32: ... C 0 3 C TP472I 5 Immersion 50 C 400 C 0 3 C 50 C 300 C 0 6 C 300 C 400 C TP472I 10 Immersion 50 C 400 C 0 3 C 50 C 300 C 0 6 C 300 C 400 C TP49A O Class A Thin Film Immersion 70 C 250 C 0 25 C TP49AC O Class A Thin Film Contact 70 C 250 C 0 25 C TP49AP O Class A Thin Film Penetration 70 C 250 C 0 25 C TP875 I Globe thermometer Ø 150 mm 30 C 120 C 0 25 C TP876 I Globe thermometer Ø 50 mm 30 C 120 ...

Page 33: ...IN Thin Film Pt100 4 wires 50 250 C 1 3 DIN TP47 1000 O 1 3 DIN Thin Film Pt1000 2 wires 50 250 C 1 3 DIN TP87 100 O 1 3 DIN Thin Film Pt100 4 wires 50 200 C 1 3 DIN TP87 1000 O 1 3 DIN Thin Film Pt1000 2 wires 50 200 C 1 3 DIN Common characteristics Resolution 0 1 C Temperature drift 20 C Pt100 0 003 C Pt1000 0 005 C ...

Page 34: ...ull scale probe 4 x 1 5V alkaline batteries operating manual case and DeltaLog9 software Other probes and the cables must be ordered separately HD2114B 0 Kit including the instrument HD2114B 0 with 600 1100mbar range barometric sensor 4 x 1 5V alkaline batteries operating manual case Other probes must be ordered separately HD2114B 2 Kit including the instrument HD2114B 2 datalogger with 600 1100mb...

Page 35: ... 2 metres TP475A 0 Air probe sensor Pt100 Stem Ø 4 mm length 230 mm Cable length 2 metres TP472I 5 Immersion probe sensor Pt100 Stem Ø 6 mm length 500 mm Cable length 2 metres TP472I 10 Immersion probe sensor Pt100 Stem Ø 6 mm length 1000 mm Cable length 2 metres TP49A O Immersion probe sensor Pt100 Stem Ø 2 7 mm length 150 mm Cable length 2 metres Aluminium handle TP49AC O Contact probe sensor Pt...

Page 36: ...Pt100 direct 4 wires Probe s stem Ø 3mm length 70mm 4 wire connection cable with connector length 2 metres TP87 1000 O Immersion probe sensor Pt1000 Probe s stem Ø 3mm length 70mm 2 wire connection cable with connector length 2 metres TP47 Only connector for probe connection Pt100 direct 3 and 4 wires Pt1000 and Ni1000 2 wires DELTA OHM metrology laboratories LAT N 124 are accredited by ACCREDIA f...

Page 37: ... 37 NOTES ...

Page 38: ... 38 NOTES ...

Page 39: ...ta da una catena di riferibilità che ha origine dalla taratura dei campioni di prima linea dei laboratori accreditati di Delta OHM presso l Istituto Primario Nazionale di Ricerca Metrologica The traceability of measures assigned to international and national reference samples is guaranteed by a reference chain which source is the calibration of Delta OHM accredited laboratories reference samples a...

Page 40: ...s of electrical and electronic equipment can return it to the dealer or manufacturer upon purchase of a new one The illegal disposal of electrical and electronic equipment is punished with an administrative fine This guarantee must be sent together with the instrument to our service centre IMPORTANT Guarantee is valid only if coupon has been correctly filled in all details Instrument code HD2114 0...

Reviews: