background image

PM[B]sense 

-   8   -

 

V1.3

 

4

 

CONFIGURATION AND MEASUREMENT 

The configuration of the instrument and the reading of the measurements can be done 
via the RS485 serial output, both with the proprietary protocol and with the MODBUS-
RTU protocol. 

In the first 10 seconds after the instrument power on, it is always active the proprie-
tary  protocol.  After  10  seconds  from  power  on,  the  operating  protocol  is  activated, 
which by default is the MODBUS-RTU protocol. 

It is possible to keep the proprietary protocol active even after 10 seconds from power on 
by sending,  before  the 10 seconds expire,  the command 

@

 of  the proprietary protocol. 

The proprietary protocol can be set as operating protocol by means of the DP0 command. 

The commands of the proprietary protocol and the registers of the MODBUS-RTU pro-
tocol are described in detail in the following chapters. 

Particulate Matter measurement modes: 

The transmitter can perform the PM measurement in continuous mode or, in order to 
extend the sensor lifetime, at cyclic intervals (default). 

In  the  measurement  mode  at  cyclic  intervals,  the  PM  sensor  is  activated  periodically 
for the ON time. The measurement is made available after 70 seconds from sensor ac-
tivation (warm-up time). At the end of the ON time, the measurement is "frozen" and 
the sensor is deactivated until the set cyclic measurement interval expires. 

 

   

Fig. 4.1: Measurement mode at cyclic intervals 

The measurement mode, the cycle interval and the sensor ON time can be configured 
respectively by using the CPLS, CPLP and CPLO commands of the proprietary protocol 
or the holding registers with address 15, 16 and 18 of the MODBUS-RTU protocol. 

The transmitter provides three types of measurements: 

 

average over a 10 seconds interval, updated every second; 

 

average over a 60 seconds interval, updated every 10 seconds; 

 

average over a 15 min interval, updated every minute. 

For  the  measurement  mode  at  cyclic  intervals,  the  cycle  interval  and  the  ON  time 
must be set consistently with the desired averaging interval. 

Status of the transmitter: 

There are two LEDs on the internal electronic board of the transmitter: the 

green

 LED 

indicates the presence of the external power supply (blinks once per second), the 

red

 

LED  indicates  the  presence  of  any  measurement  errors  (normally  off,  it  blinks  twice 
per second if at least one of the measured parameters is in error). 

ON time 

Cycle interval 

Sensor ON 

Sensor OFF 

Summary of Contents for PMBsense-A

Page 1: ...Operating manual PM CO2 Transmitter PM B sense www deltaohm com English Keep for future reference...

Page 2: ...S 1 INTRODUCTION 3 2 TECHNICAL SPECIFICATIONS 3 3 INSTALLATION 5 3 1 ELECTRICAL CONNECTIONS 5 4 CONFIGURATION AND MEASUREMENT 8 5 PROPRIETARY PROTOCOL 9 6 MODBUS RTU PROTOCOL 15 7 SAFETY INSTRUCTIONS...

Page 3: ...on request when order ing analog outputs is available The two analog outputs can be independently associ ated with any of the measured parameters The measuring circuit of the transmitter can be operat...

Page 4: ...esponse time 120 s air speed 2 m s Long term stability 5 of measurement 5 years Temperature drift 1 ppm C General specifications Output RS485 with Modbus RTU or ASCII proprietary protocol Only PMsense...

Page 5: ...escription 3 1 ELECTRICAL CONNECTIONS The transmitter has a M12 8 pole circular connector and uses the CPM12 8PM op tional cables with M12 8 pole connector on one side and open wires on the other side...

Page 6: ...are short circuited inside the transmitter The cable shield is not connected to the M12 connector To ensure a good noise immunity it is recommended to connect the cable shield to ground GND RS485 conn...

Page 7: ...OUT1 GND Vdc AOUT2 Fig 3 4 current analog outputs connection Voltage analog output optional only PM B sense A By default the voltage analog outputs are 0 10 V with 0 V 0 g m3 10 V 1000 g m3 The analog...

Page 8: ...riodically for the ON time The measurement is made available after 70 seconds from sensor ac tivation warm up time At the end of the ON time the measurement is frozen and the sensor is deactivated unt...

Page 9: ...USER ON must be sent first the instrument replies with USER CAL MODE ON The command CAL USER ON is automatically disabled after a few minutes of inactivity If the settings should be only read the com...

Page 10: ...ting of parity and stop bits CMWn Sets waiting time after transmission with MODBUS RTU protocol Immediate reception if n 0 violates protocol Waiting 3 5 characters if n 1 respects protocol Default Imm...

Page 11: ...second the sequence is the same described in command P5 P5 Measurements Prints the PM measurements in the following sequence PM measurement error 0 no 1 yes PM1 0 in N of particles ml PM2 5 in N of pa...

Page 12: ...register 19 RA1T n Reads the physical quantity associated to analog output 1 CA1Ln Sets n as the minimum value of the measuring range of the physical quantity associated to the analog output 1 If the...

Page 13: ...n Reads the physical quantity associated to analog output 2 CA2Ln Sets n as the minimum value of the measuring range of the physical quantity associated to the analog output 2 If the output is associ...

Page 14: ...ter must first be placed in an environment with a known CO2 concentration the value n entered in the command The value t in the reply to the command indicates the number of seconds required for the tr...

Page 15: ...ed according to what is set with CPS command or MODBUS holding register 19 16 bit Integer 2 PM10 in N of particles ml Measurement averaged according to what is set with CPS command or MODBUS holding r...

Page 16: ...every mi nute 16 bit Integer 20 PM10 in N of particles ml Measurement averaged over 15 min and updated every mi nute 16 bit Integer 21 PM1 0 in g m3 x10 Measurement averaged over 15 min and updated ev...

Page 17: ...set disabled 0 20 mA or 0 10 V default if the out put is voltage 1 offset enabled 4 20 mA or 2 10 V default if the output is current Bit 4 Sets the direct or inverse correspondence between analog outp...

Page 18: ...cal quantity associated to the analog output 1 If the output is associated to PM the value is expressed as a number of tenths e g 5 0 5 g m3 If the output is asso ciated to CO2 the value is expressed...

Page 19: ...val in seconds for the PM meas urement mode at cyclic intervals The default value is 300 5 min 16 bit Integer Unsigned 18 Setting of the sensor ON time in seconds for the PM measurement mode at cyclic...

Page 20: ...well as the specific measures described in this manual are followed The instrument proper operation and operating safety can be ensured only in the cli matic conditions specified in this manual Do not...

Page 21: ...8PM 5 or 10 m CPM12 8PM 10 standard other lengths on request RS51K Kit for connecting the transmitter RS485 output to a PC It includes the SWD10 power supply and the RS485 USB adapter with o screw ter...

Page 22: ...PM B sense 22 V1 3 NOTES...

Page 23: ......

Page 24: ...rranty included in the manual of the product For any dispute the competent court is the Court of Padua The Italian law and the Convention on Contracts for the International Sales of Goods apply TECHNI...

Reviews: