PM[B]sense
- 11 -
V1.3
Reading of the measurement information:
Command
Reply
Description
CPSn
&|
Sets the type of PM measurement averaging for the ana-
log outputs and the measurements sent by the transmitter
in reply to P1, P5, S1 and S5 commands or when reading
MODBUS Input Registers 0…5:
Average over a 10 seconds interval, updated every
second if n=0
Average over a 60 seconds interval, updated every
10 seconds if n=1
Average over a 15 minutes interval, updated every
minute if n=2
Default
: Average over a 10 seconds interval, updated
every second (n=0)
RPS
& n|
Reads the type of averaging for the measurements sent by
the transmitter in reply to P1, P5, S1 and S5 commands.
P0
&|
Disable the sending of the measurement enabled with P1.
P1
&|
Enable the sending of the PM measurement every second
(the sequence is the same described in command P5).
P5
&
Measurements
Prints the PM measurements in the following sequence:
PM measurement error (0=no, 1=yes)
PM1.0 in N° of particles/ml
PM2.5 in N° of particles/ml
PM10 in N° of particles/ml
PM1.0 in µg/m
3
PM2.5 in µg/m
3
PM10 in µg/m
3
The PM measurement is averaged according to what is set
with CPS command or MODBUS holding register 19.
S0
&|
Disable the sending of the measurement enabled with S1.
S1
&|
Enable the sending of the measurement every second
(the sequence is the same described in command S5).
S5
&
Measurements
Prints the measurements in the following sequence:
PM measurement error (0=no, 1=yes)
PM1.0 in µg/m
3
PM2.5 in µg/m
3
PM10 in µg/m
3
CO
2
in ppm
Atmospheric pressure in hPa (internal sensor for
CO
2
measurement compensation)
Field not used
Field not used
Field not used
Power supply voltage
Internal board temperature
The PM measurement is averaged according to what is set
with CPS command or MODBUS holding register 19.