PM[B]sense
- 10 -
V1.3
Note
: after sending the DP1 command, the instrument remains with the proprietary protocol.
Send the command SM to activate the MODBUS-RTU protocol immediately, or power cycle the
instrument.
RS485 communication parameters:
Command
Reply
Description
CMBn
&|
Sets the Baud Rate:
1200 if n=0
▪
19200 if n=4
2400 if n=1
▪
38400 if n=5
4800 if n=2
▪
57600 if n=6
9600 if n=3
▪
115200 if n=7
Default
: 19200 (n=4)
RMB
& n|
Reads Baud Rate setting
CMPn
&|
Sets parity and stop bits:
8N1 if n=0 [No parity, 1 stop bit]
8N2 if n=1 [No parity, 2 stop bits]
8E1 if n=2 [Even parity, 1 stop bit]
8E2 if n=3 [Even parity, 2 stop bits]
8O1 if n=4 [Odd parity, 1 stop bit]
8O2 if n=5 [Odd parity, 2 stop bits]
The number of data bits is fixed to 8.
Default
: 8E1 (n=2)
RMP
& n|
Reads the setting of parity and stop bits.
CMWn
&|
Sets waiting time after transmission with MODBUS-RTU
protocol:
Immediate reception if n=0 (violates protocol)
Waiting 3.5 characters if n=1 (respects protocol)
Default
: Immediate reception (n=0)
RMW
& n|
Reads the setting of waiting time after transmission with
MODBUS-RTU protocol.
PM measurement settings:
Command
Reply
Description
CPLSn
&|
Sets the PM measurement mode:
Continuous if n=0
At cyclic intervals if n=1
Default
: At cyclic intervals (n=1)
RPLS
& n|
Reads the setting of the PM measurement mode.
CPLPn
&|
Sets the cycle interval for the measurement mode at cy-
clic intervals.
Default
: 300 (=5 min)
RPLP
& n|
Reads the setting of the cycle interval for the measure-
ment mode at cyclic intervals.
CPLOn
&|
Sets the sensor ON time for the measurement mode at
cyclic intervals to n seconds.
It must be greater than 70 s (warm-up time).
Default
: 71
RPLO
& n|
Reads the setting of the sensor ON time for the measure-
ment mode at cyclic intervals.
Summary of Contents for PMBsense-A
Page 22: ...PM B sense 22 V1 3 NOTES...
Page 23: ......