19
COMMAND
ACTION
RESPONSE
NOTES
KX
Start measurement
&
LDxxxx
Send sample
no. xxxx+1
Dump or ?
Xxxx from 1 to 1000
LE
Clear memory
&
LN
No. of next memory location
request
Next avail. Memory =0001
MR
Type of measurement reading
&n
n = see MW command
MW0..5
Type of measurement selection
0 = EPA 180
1 = ISO NEPH
2 = EBC
3 = ASBC
4 = WHITE %
5 = IR %
P0
Ping & lock keys
&
P1
Ping & unlock keys
&
RH
Shows the calibration
validity number of days
RI
Sample
ID=
00000001
RL
Print&mem = 0
0 = free print,
1 = print & store,
(see the MEM ON PRNT
command in the menu)
RP
Battery voltage reading
&nnn
Nnn = hundredths of Volt
SH
Calibration status reading
calibration status = valid
calibration status = expired!
T3Do35
User calibration reset, return to
factory calibration
&
T4Do35nnnnn
xxxxx
Enable user to change its
password
&/?
Nnnnn = current administrator
password
Xxxxx = user password being
changed
WI
Enter sample identification
number
&/?
00000000 ... 00019999
WL
Choice of print-memory
connection mode
&/?
0 = free print,
1 = print and forced storing
WUnnnnn
Set new password of user
previously selected with
command T4Do35
&/?
Command characters are exclusively upper case characters. Once a correct command is entered, the
instrument responds with "&"; when any wrong combination of characters is entered, the instrument
responds with "?". The instrument response strings end with the sending of the “|”+CR+LF
command (| + Carriage Line Feed). Before sending commands to the instrument via the
serial port, locking the keyboard to avoid functioning conflicts is recommended: use the P0
command. When complete, restore the keyboard with the P1 command.