Test A: 

 

 

Attach the #831

 short pin prod to the H-4 handle and connect the handle to the 

external connector on top of the meter. 

 

 

Prepare a representative sample

 by collecting hay from various parts of the 

windrow.   

 

 

Place hay in a non-conductive container

 (such as a 5 to 10 gallon plastic pail) and 

apply the short pin prod to the hay. 

 

 

Press the read button (#1) 

and take a reading. 

 

 

Mix the sample once again

 and take at least two more readings.  Use the highest 

readings.  

 
NOTES 
 

 

Whenever pressure is being applied on the electrode, be sure that the points of the 
electrode are touching nothing but the hay. 

 

 

Repeat the steps above if considerable variations are found in the meter readings. To 
reduce these variations, chop the hay, mix it thoroughly and take several readings by 
following the procedures above. This will make the moisture distribution in the sample 
more uniform 

 
Test B: 

 

 

Attach the #831 short pin prod

 to the H-4 handle and connect the handle to the 

external connector on top of the meter. 

 

 

Apply the prod

 to the hay in the windrow. 

 

 

Press the read button #1 

and take a reading. 

 

 

Make several tests

 on the hay exposed to the sun, then turn the windrow over and 

make an equal number of tests on the hay that had been closer to the ground.  Use the 
highest readings. 

 
NOTES 

  

 

Make sure that the points of the electrode are not touching the ground.  The electrode 
points should make contact with the hay only. 

 

Test C: 

 

 

Select up to five large, slower-drying stems

 from a section of the windrow. 

 

 

Place them one at a time 

across two adjacent points on the #831 short pin prod. 

 

 

The average of these stem readings

 should be about two to five points higher than 

the actual moisture content. 

 
NOTES 
 

 

Repeat these steps in different parts of the field and pay special attention to the areas 
where the hay is heaviest. 

 

Summary of Contents for F-2000H

Page 1: ...F 2000H HAY AND HOPS MOISTURE METER OPERATION MANUAL v3 6 DELMHORST EUROPE TITANIUMLAAN 100 NL 5221 CK s HERTOGENBOSCH THE NETHERLANDS info delmhorst eu www MoistureMetersDelmhorst com 31 0 73 6395080...

Page 2: ...Testing Baled Hay 5 Testing in the Windrow 7 Factors Affecting Your Reading 7 Range Of Moisture Content 7 Hay Temperature 8 Curing 8 Density 8 Use Of Preservatives 8 Sample Size 9 Testing for hops 10...

Page 3: ...ing erases the readings Scale Button 3 Displays the meter scale hay or hops and acts as a toggle switch to change between two scales Also acts as an arrow button when pressed after the set point butto...

Page 4: ...value To scroll forward to a higher value hold the set point button 4 while the current value is displayed and scroll to the set point value desired To scroll backward through the set point values pre...

Page 5: ...tored in memory TAKING A READING TESTING BALED HAY Connect the probe to the external connector on the top of the meter Insert the probe into the bale Press the read button 1 The meter displays the MC...

Page 6: ...ve This will make the moisture distribution in the sample more uniform Test B Attach the 831 short pin prod to the H 4 handle and connect the handle to the external connector on top of the meter Apply...

Page 7: ...creases so does the conductivity Tests on hay at high moisture content over 25 are less accurate This is mostly due to the variability in moisture distribution The reduced level of accuracy in the hig...

Page 8: ...ll ensure firmer and more uniform contact USE OF PRESERVATIVES Hay preservative or stabilizers may also have an effect on meter readings Normally a bale of hay treated with preservative will read high...

Page 9: ...the moisture level in the hops that is in contact with the uninsulated tip of the prod Therefore it is very important to make a number of tests in all parts of a bale for proper testing Since moistur...

Page 10: ...numbers or e mail address If you believe the meter is under warranty please provide the original sales slip or invoice Ship via UPS Express Mail Priority Mail or any overnight courier who provides pro...

Page 11: ...MHORST be liable for any incidental indirect special or consequential damages of any type whatsoever including but not limited to lost profits or downtime arising out of or related in any respect to t...

Page 12: ...ective Standard s to which Conformity is declared EN 50082 1 1992 EMC Generic immunity standard EN 50081 1 1992 EMC Generic emission standard EN 55011 1991 Limits and methods of measurement of radio d...

Reviews: