To erase all the accumulated readings

, hold the calibration check button (#2) for 

more than five seconds until the meter displays “0.” 

 

To keep the accumulated readings in memory, 

release the calibration check button 

(#2) before the total cycle time is complete.  

 
The meter will accumulate up to 100 readings.  After all 100 readings are stored, it will not 
add new readings until the memory has been cleared.  It will also continue to display the 
average of all 100 readings as a reminder that the memory is full. 
 

Readings below 6% will be displayed as “0”.  Those above 40% will be displayed as 
“99.9.”  Neither will be added to the accumulated readings or used in calculation of 
average or highest reading. 

 
 
TO RESET THE METER 
 

 

Press and release

 the calibration check button (#2).   

 

 

Within one second

, press and hold the scale button (#3). 

 

 

The meter will reset itself

 and display “119.” This indicates that the meter has been 

reset to scale #1 (hay) and default set-point 19%.  It will also clear all of the readings 
stored in memory. 
 

 

TAKING A READING 

 
TESTING BALED HAY 
 

 

Connect the probe

 to the external connector on the top of the meter. 

 

 

Insert the probe into the bale

 

 

Press the read button (#1).

  The meter displays the %MC for two seconds. 

 
NOTES

 

 

 

The hay prod is electrically insulated, except at the metal points near the tip.  The 
moisture content measured represents the hay in contact with the tip of the prod only.   

 

 

Partially cured hay may have wide variations in moisture content throughout the bale. 
Readings should be taken in several different parts of the bale and the highest readings 
used as a guideline.  The arrangement and compaction of hay fibers in a bale may have 
an effect on meter readings. 

 

 

If you are testing high density bales, we recommend using the H-4 handle with the 
830-2 25cm prod, 830-3 45cm prod, or the 830-4 90cm prod.  Using the handle/prod 
combination eliminates excess stress on the instrument case that may occur when trying 
to insert the prod into a high density or large bale. 

 

 

When using the 90cm prod, be sure to guide the prod into the bale with one hand while 
pushing on the H-4 handle.   

 

TESTING IN THE WINDROW 

 
There are three ways to test moisture content in the windrow: 
 

 

 

Summary of Contents for F-2000H

Page 1: ...F 2000H HAY AND HOPS MOISTURE METER OPERATION MANUAL v3 6 DELMHORST EUROPE TITANIUMLAAN 100 NL 5221 CK s HERTOGENBOSCH THE NETHERLANDS info delmhorst eu www MoistureMetersDelmhorst com 31 0 73 6395080...

Page 2: ...Testing Baled Hay 5 Testing in the Windrow 7 Factors Affecting Your Reading 7 Range Of Moisture Content 7 Hay Temperature 8 Curing 8 Density 8 Use Of Preservatives 8 Sample Size 9 Testing for hops 10...

Page 3: ...ing erases the readings Scale Button 3 Displays the meter scale hay or hops and acts as a toggle switch to change between two scales Also acts as an arrow button when pressed after the set point butto...

Page 4: ...value To scroll forward to a higher value hold the set point button 4 while the current value is displayed and scroll to the set point value desired To scroll backward through the set point values pre...

Page 5: ...tored in memory TAKING A READING TESTING BALED HAY Connect the probe to the external connector on the top of the meter Insert the probe into the bale Press the read button 1 The meter displays the MC...

Page 6: ...ve This will make the moisture distribution in the sample more uniform Test B Attach the 831 short pin prod to the H 4 handle and connect the handle to the external connector on top of the meter Apply...

Page 7: ...creases so does the conductivity Tests on hay at high moisture content over 25 are less accurate This is mostly due to the variability in moisture distribution The reduced level of accuracy in the hig...

Page 8: ...ll ensure firmer and more uniform contact USE OF PRESERVATIVES Hay preservative or stabilizers may also have an effect on meter readings Normally a bale of hay treated with preservative will read high...

Page 9: ...the moisture level in the hops that is in contact with the uninsulated tip of the prod Therefore it is very important to make a number of tests in all parts of a bale for proper testing Since moistur...

Page 10: ...numbers or e mail address If you believe the meter is under warranty please provide the original sales slip or invoice Ship via UPS Express Mail Priority Mail or any overnight courier who provides pro...

Page 11: ...MHORST be liable for any incidental indirect special or consequential damages of any type whatsoever including but not limited to lost profits or downtime arising out of or related in any respect to t...

Page 12: ...ective Standard s to which Conformity is declared EN 50082 1 1992 EMC Generic immunity standard EN 50081 1 1992 EMC Generic emission standard EN 55011 1991 Limits and methods of measurement of radio d...

Reviews: