Changing the password
Changing the password
Setting the year
Setting the year
Setting the date
Setting the date
Setting the time
Setting the time
Exit
Calibration
a)
Connect fresh air to the sensor input. The icon
„
“
flashes
.
After the value stabilizes, move onto the next step using
„ “
b)
Using „ “ select the concentration of the calibration gas, confirm
„ “
c)
Connect the calibration gas to the sensor input. The icon
„ “
flashes
. Wait until the value stabilizes and confirm
„ “
d)
Using „ “ swtch between „YES“ - save calibration, or „NO“ - return to the Settings menu. Confirm the selected
option „ “
e)
In case of saving a new calibration, the transmitter will restart.
Setting the alarms
a)
Using „ “ select the concentration for alarm 1. Confirm „ “
b)
Using „ “ select the concentration for alarm 2. Confirm „ “
c)
Using „ “ select the concentration for alarm 3. Confirm „ “
d)
Using „ “ select the concentration for alarm 4 Confirm „ “
e)
Using „ “ switch between „YES“ - save settings, or „NO“ - return to the Settings menu. Confirm the selected
option „ “
Setting the range of the 20mA current loop
a)
Using „ “ select the concentration appropriate to 20mA range. Confirm „ “
b)
Using „ “ switch between „YES“ - save settings, nebo „NO“ - return to the Settings menu. Confirm the selected
option „ “
Changing the password
a)
Using „ “ select a number in thousands. Save the selected number „ “
b)
Using „ “ select a number in hundreds. Save the selected number „ “
c)
Using „ “ select a number in tens. Save the selected number „ “
d)
Using „ “ select a number in units. Save the selected number „ “