20
contact the
manufacturer
13
Error reading the internal
FLASH
Restart the transmitter.
If the error persists,
contact the
manufacturer
14
Error reading the internal
FLASH
Restart the transmitter.
If the error persists,
contact the
manufacturer
15
Sensor error
Replace the sensor
16
The device exceeded the
maximum calibration
interval by 50%
Calibration is necceserry
129
The measured value
exceeds the range of -
10%
If the error persists,
contact the
manufacturer
130
The result of the AD
covertion is out of range
Contact the
manufacturer
3. Signalization transmitted by the current loop 4-20mA
Measurement:
The measure concentration is directly proportional to 4-20mA current output
Exceeding the range of measured concentrations:
Current output ranges from 20-22mA
End of valid calibration:
Current output transmits the actual measured concentration for 9s in a 4-20mA range and for 1s 2mA
current
Malfunction:
Current output will be set to 0,5mA
Mainternace:
Current output will be set to 1mA