6
Auto Ign. 20/90
Test Time
The
PosiTector
SHD
has an on-screen timer to measure the
hardness of a material after a given period of time as defined by
international standards, typically 1, 3 or 15 seconds.
Use the
and
buttons to adjust the test duration. When
taking a reading, the timer starts automatically once the presser
foot is on the surface and the ▼ symbol appears on the display.
According to most international standards, readings below 20
and above 90 are not considered reliable. When enabled,
measurement results outside 20/90 will emit a low tone and will
not be included in
Memory
or
Statistics
.
If values above 90 Shore A are measured, the
PosiTector
SHD D
probe should be used. If values below 20
Shore D are measured, use the
PosiTector
SHD A
probe.
NOTE:
(Auto Ignore)
Adjusts the volume of built-in speaker (Off, Low, Medium, High).
Sound
Disables the
Auto Rotate
feature by locking the display in its
current orientation.
Flip Lock
Allows the touch screen functionality to be disabled. All gage
functions can also be controlled using the navigation buttons.
Touch
All measurements are date and time stamped (24-hour format)
when stored into memory. It is therefore important to set the
correct date and time. Use the
and
buttons to select a
value, and the
and
buttons to adjust it. The current date
and time setting can also be viewed at the top of the main menu.
Set Clock