Table 4.2:
Connectors for cables on the front plane of the detector electronics unit
Connectors
Description
DATA1
Data cable to the vacuum feedthrough / detector head (blue
Samtec SQCD ribbon cable)
POWER1
Power cable to the vacuum feedthrough / detector head (grey
Sub-D cable)
Table 4.3:
Status LEDs on the detector back plane
LED
Description
LINK
Orange if detector and PC are powered and connected via the
RJ45 data cable.
ACT
Flashing green, if there is activity on the data line between detector
and DCU.
EN
Orange, if the detector is in counting mode.
TEMP
Normally green. Turns red if the detector temperature or humidity
is out of the limits.
POWER
Normally green. Turns red if there is a power failure or if the de-
tector temperature or humidity is out of the limits.
Table 4.4:
Connectors for cables on the back plane of detector electronics unit
Connectors
Description
DATA
RJ45 Cat 6A S/FTP cable for data transfer.
Caution
#6
Note that there must be a 1 Gbit point-to-point con-
nection between detector and PC (Gb2).
+12 V DC
Main voltage 12 V DC from the external power supply
EXT IN
External trigger input. Use a Lemo
®
Type 00 (NIM/CAMAC)
cable.
EN OUT
TTL Level (5 V) output signal; high when counting is enabled.
Use a Lemo
®
Type 00 (NIM/CAMAC) cable.
Functional ground
Information
#4
Although the detector might be already grounded via
the mounting bolts, the detector should be grounded
additionally via the functional ground connector at the
back to establish a defined grounding.
FUSE 4 AT
Fuse (see table 3.3)
PILATUS3 R CdTe 100K-M Technical Specifications v1.0.0
14 | 29