SERIES 5 SIDE DELIVERY SYSTEM

Instructions For Use

MAINTENANCE

SELF-CONTAINED WATER SYSTEM

The self-contained water system is designed to optimize the quality of water being delivered to the handpieces and 
syringes.  The self-contained water system has three functions:

1. 

The system may be filled with filtered or sterile water for patient use.

2. 

The system may be filled with disinfectant for flushing the syringe and handpiece tubing.

3. 

The unit can be purged with air to inhibit the growth of biofilm.

In order to maintain a truly aseptic water system, daily and weekly cleaning procedures must be performed in a consistent, 
regular manner.  Failure to do so could result in contaminated water lines and a lower water quality than what is acceptable 
for patient use.  Follow a daily and weekly maintenance routine for proper cleaning of the tubing with the clean water 
system.

Fill bottle with the 100 ml disinfectant solution, shake vigorously and let it settle for 10 minutes.  Shake again, then rinse 
twice with water.

It is recommended that 100 ml of disinfectant solution is mixed for each weekly bottle disinfecting procedure.  Always use a 
fresh mixture every week.

The Disinfectant Solution:  9 parts (90 ml) Tap water

1 part (10 ml) 5.25% Sodium hypochlorite (household bleach)

CAUTION:  

Disinfect new water bottle prior to first use.

DISINFECTING THE BOTTLE

Summary of Contents for DW5050

Page 1: ...SERIES 5 SIDE DELIVERY SYSTEM Instructions For Use...

Page 2: ...2 OVERVIEW 2 SERIES 5 SIDE DELIVERY SYSTEM Instructions For Use SIDE DELIVERY SYSTEM DW5050 DW5550...

Page 3: ...TENTS 2 3 4 5 6 7 8 9 10 14 15 18 WARNING Failure to carefully follow the described procedure may result in damage to the equipment the operatory and or the operator NOTE Take note of additional impor...

Page 4: ...ws as a regular maintenance item to ensure proper positioning of the device IEC Medical Classification 2A TYPE BF Power optics Splash Protection IPX0 TECHNICAL DOCUMENTATION The manufacturer will make...

Page 5: ...uted without increase risk of shock or fire Use authorized replacement parts only Power cords must be installed by qualified personnel Ensure all service loops strain reliefs and cord guards are in pl...

Page 6: ...YSTEM Instructions For Use Unit Keypad Function List Key Description Initiate Function Cancel Function Press key to intiate function Press any key to cancel function Release key to cancel function Pre...

Page 7: ...7 OPERATION CONTROL HEAD 7 SERIES 5 SIDE DELIVERY SYSTEM Instructions For Use CONTROLS...

Page 8: ...ieces Rotate counterclockwise to increase flow clockwise to decrease flow HANDPIECE FLOW CONTROLS Located at the front of the control head these individual control valves adjust the amount of water co...

Page 9: ...e is observed turn the Water Bottle On Off switch OFF to release all pressure before tightening the bottle to stop leakage 2 Select either city water or bottled water supply source as desired using th...

Page 10: ...rclockwise to increase pressure DRIVE AIR FLOW ADJUSTMENT DRIVE AIR PRESSURE GAUGE UTILITY CENTER The Utility Center comes factory preset at 40 psi for Water pressure and 80 psi for Air pressure All r...

Page 11: ...inge to prevent splashing and to achieve a desirable mist pattern The adjusting screws are located under the unit head directly below the syringe control block SYRINGE CONTROL BLOCK ADJUSTMENT SCREWS...

Page 12: ...nuckle B 5 When the unit head is level retighten the set screw in the stop block to secure the adjustment A HEAD LEVELING FRONT FLEX ARM AND HEAD ROTATION Rotational tension is preset at the factory h...

Page 13: ...knuckle clockwise with a 3 16 hex key B Retighten the knuckle block set screw A Replace the end caps TO DECREASE SPRING TENSION 1 Remove the end cap at the front of the flex arm and loosen the set scr...

Page 14: ...es at their recommended strengths can be corrosive at higher than recommended dilution ratios CHEMICAL DISINFECTION CAUTION These disinfectants will harm the surface finish of dental equipment and are...

Page 15: ...rovide additional recommendations regarding this procedure including information on frequency and duration of flushing and the use of antibacterial solutions in the self contained water system MAINTEN...

Page 16: ...stem again until all the cleaning solution is used up 4 Purge the unit with air a Hold the handpiece tubing and syringe over a container Turn the Master Switch to the On position wait a few moments th...

Page 17: ...cleaning procedures must be performed in a consistent regular manner Failure to do so could result in contaminated water lines and a lower water quality than what is acceptable for patient use Follow...

Page 18: ...ld be worn to reduce the possibility of damage ACCESSORY USE Using accessory devices not specified by DCI for use with their equipment may result in an increase of electromag netic emissions and or a...

Page 19: ...DCI International LLC 305 N Springbrook Road Newberg OR 97132 USA PN 50196 Rev A 2021 08 23 www dcionline com 800 624 2793...

Reviews: