SERIES 5 SIDE DELIVERY SYSTEM
Instructions For Use
ADJUSTMENTS
1. Use a 3/32” hex key or the ball driver provided
with the unit to make the syringe flow adjustment.
Identify which adjusting screw is for air and which
is for water by the color of the tubing connected
to the block. Blue is water, yellow is air.
2.
Adjust the water first, with the syringe button fully
depressed. Turn the screw clockwise to decrease
flow or counterclockwise to increase flow.
3.
After adjusting the water to the desired flow, press
both buttons simultaneously and adjust the air flow
to achieve a mist pattern that suits your needs.
DOCTOR’S SYRINGE
Adjusting screws allow you to control the flow of air and water from the syringe to prevent splashing and to achieve a
desirable mist pattern. The adjusting screws are located under the unit head directly below the syringe control block.
SYRINGE CONTROL BLOCK
ADJUSTMENT SCREWS
The rotational tension of the flex arm may be adjusted at the set screw on the post, with a 3/32” hex key, located below
the joint where the rear knuckle of the flex arm joins the rigid arm. Similarly the rotational tension of the rigid arm can be
adjusted via the set screw on the opposite end of the rigid arm.
REAR FLEX ARM AND RIGID ARM ROTATION
HOLDER
The handpiece holders come pre-spaced and leveled from the factory, but
may be re-positioned for the user’s needs.
1. With a 1/8” hex key, loosen the two set screws located on the
bottom of the holder.
2. Reposition the holder in the desired location.
3. Retighten the two set screws with the 1/8” hex key to secure the
holder in position.
SET SCREWS
SET SCREWS