The Model 5M78 is a single-channel conditioner of phase-sensitive carrier-amplifier
design.  Intended for applications involving transformer-coupling to the transducer 
bridge (as with rotary-transformer torque sensors), this conditioner can also be used in
conventional installations when high sensitivity is required or where the electrical 
environment is especially noisy. Responding only to the modulated carrier frequency, 
the 5M78 rejects extraneous voltages that can cause errors in DC systems,
particularly when there is a need to “blow up” a portion of the transducer range. 
User-adjustable phase and symmetry controls are provided.

calibration resistor is 59 kilohms, 1% which is located internally to the 5M78 case.

1

G

ENERAL

D

ESCRIPTION AND

S

PECIFICATIONS

Model 5M78 AC Strain Gage Module

Access switch settings via the front

panel of the 5M78 by gently pulling the 
clear plastic cover (from the bottom 
side) so the cover rotates open from 

DIN AC Strain Gage  Conditioner

M

ODEL 

5M78

AC Strain Gage

C

ONDITIONER 

Module

5M78

 

  

1.A.1.5M78

G

ENERAL 

D

ESCRIPTION AND 

S

PECIFICATIONS 

5M78.1

or without power to the unit. Once 

completed, return the cover to the 

View of Side Label of the Model 5M78 AC Strain Gage Module

the top. Use a small tool or finger to 

place the switches to the left or right 
position as you face the front of the 
module. This process can be done with 

original position. 

AC Strain Gage Phase Operation

Due to the AC modulated aspects of the 

5M78 AC Voltage excitation circuit, the 5M78
contains a "PHASE" adjustment which 
aligns the transducer's return output signal  
to the conditioner's demodulator. When 

the  demodulator is aligned properly through 
calibration, maximum amplitude and 
accuracy are achieved. 

The Model 5M78's is calibrated by means of a “two-point (dead-weight)” or 
 shunt-calibration technique, which are outlined in section three.  The supplied shunt

92342.00

 

Summary of Contents for 5M78

Page 1: ...the 5M78 by gently pulling the clear plastic cover from the bottom side so the cover rotates open from DIN AC Strain GageConditioner MODEL 5M78 AC Strain Gage CONDITIONER Module 5M78 1 A 1 5M78 GENERA...

Page 2: ...miento de este instrumento debe ser realizada por personal calificado solamente El C digo El ctrico Nacional define a una persona calificada como una que est familiarizada con la construcci n y operac...

Page 3: ...r elektrische Sicherheitsauflagen f r Angestellt Arbeitspl tze ANSI NFPA 70E 2012 der Vereinigten Staaten und alle zus tzlichen Arbeitsplatzsicherheitsauflagen folgen die auf Ihre Installation anwend...

Page 4: ...y que el material aislante no tenga rajaduras Reemplace de inmediato cualquier parte defectuosa Mesures de S curit Les mesures de s curit suivantes doivent tre prises chaque fois qu un type de connex...

Page 5: ...achenden Schaltung hergestellt werden Tragen Sie Schutzbrillen und Isolierhandschuhe wenn Sie Anschl sse an den Stromkreisen vornehmen H nde Schuhe und Fu boden m ssen trocken sein wenn Sie Anschl sse...

Page 6: ...al value 0 08 sec 0 008 sec 100 Hz 0 0011 sec 650 Hz To 0 1 of final value 0 095 sec 0 0095 sec 100 Hz 0 0016 sec 650 Hz To 0 02 of final value 0 10 sec 0 01 sec 100 Hz 0 0014 sec 650 Hz 5M78 1 GENERA...

Page 7: ...SENSE Top Front 4 4 SHIELD Bottom Front 1 1 Ano Out ANALOG Output Bottom Front 2 2 SHP Shunt Positive Bottom Front 3 3 SHN Shunt Negative n Bottom Front 4 4 SYC SYNC MASTER SLAVE TRANSDUCER CONNECTION...

Page 8: ...ration resistor for convenience in subsequent checking Replace the installed calibration resistor with one or an equivalent resistance value supplied by the transducer manufacturer to achieve a precis...

Page 9: ...ent dead weight value greater than one half of full scale Adjust Coarse Span for a nominal corresponding output Adjust the PHASE control until a maximum output value is obtained Once set for your tran...

Page 10: ...ted on the calibration document The SYMMETRY control has approximately 7 authority full scale If dead weight calibration is not practical and the transducer calibration data is unknown the Equivalent...

Page 11: ...Switch 2 Volts selects 5 or 10 Vdc when mode is voltage Switch 3 Current selects 4 12 20 or 4 20mA when mode is current Switch 4 Filter selects 650 Hz or 100 Hz at 3 db for 100 Hz switch 5 must be se...

Page 12: ...made is determined by the direction of the loading or off zero reading c The inability to balance correctly where the instrument output reads totally off scale and the BALANCE controls have no authori...

Page 13: ...Contact Switch NO NOT PUSHED Logic 1 Shunt not enabled PUSHED Logic 0 Shunt is enabled Push Button NO 11 28 VDC POWER SUPPLY SHIELD Fig 3 Shunt Connections Fig 4 Star Bridge...

Page 14: ...urment setup beat frequencies can be produced by the excitation clock oscillator To prevent this the user should declare one of the 5M78 units as a Master via switch 6 on the front panel The other uni...

Page 15: ...orkmanship or the refund of the purchase price to the Buyer Before returning a product or products for any reason the customer must call Daytronic Customer Support Services at ALL EQUIPMENT TO BE REPA...

Page 16: ...Daytronic Corporation Dayton OH USA w ww ww w d da ay yt tr ro on ni ic c c co om m...

Reviews: