USER’S MANUAL
MA07-02/B
16
6. TROUBLESHOOTING
Error Indicator
Error Indicator
Error Indicator
Error Indicator
Cause
Cause
Cause
Cause
Correction
Correction
Correction
Correction
WEIGHT ERROR
CODE 01
Error weight reading
-
Failure in load cell
-
Failure in A/D
-
Failure in load cell wires
OVERFLOW
CODE 02
Weight over 225 kg
-
Failure in load cell
-
Failure in A/D
-
Failure in load cell wires
UNDERFLOW
CODE 03
Weight under 0 kg
-
Failure in load cell
-
Failure in A/D
-
Failure in load cell wires
-
Failure in load cell mounting
ZERO WEIGHT ERROR
CODE 04
Error in Zero reading at
machine start.
-
Failure in load cell
-
Failure in A/D
-
Failure in load cell wires
-
Failure in load cell mounting
-
Knocks in load cell
HEIGHT ERROR
CODE 05
Height reading error.
-
Failure in height cell.
-
Failure in height circuit.
-
Failure in height wires.
-
Failure in power 12V
PNEUMATIC ERROR
CODE 10
Error on the E2Prom
memory.
-
Check if the
Pressure electronic board is
well inserted.
-
Replace the
Pressure electronic board.
PNEUMATIC ERROR
CODE 11
Cuff over-inflated. More than
300 mmHg.
-
Failure in air pressure.
-
Failure in pressure measuring.
PNEUMATIC ERROR
CODE 12
The inflating process time
has exceeded.
-
Failure in air pressure circuit.
-
Failure in pressure measuring.
PNEUMATIC ERROR
CODE 13
Problem caused by abnormal
function during the detection
process.
-
Failure in air pressure circuit.
-
Failure in pressure measuring.
PNEUMATIC ERROR
CODE 14
Problem caused by abnormal
memory function.
-
Failure en la E2PROM de la PressV3.
-
Failure in I2C circuit.
PNEUMATIC ERROR
CODE 15
The process has been
cancelled by STOP button.
-
User push STOP keypad to abort
pressure measurement.
PNEUMATIC ERROR
CODE 16
The measuring process time
has exceeded.
-
Failure in air pressure circuit.
-
Failure in pressure measuring.
PNEUMATIC ERROR
CODE 17
No weight detection on the
platform.
-
To measure the blood pressure you
must stand on the platform.
PNEUMATIC ERROR
CODE 18
Blood pressure measuring
error.
-
Data type very difficult to analyze : low
signal, abnormal pressure with
arrythmics, etc
BODYFAT ERROR
CODE 20
Reading error in Bodyfat
-
Comms error in I2C
-
Reset failure in Bodyfat PCB
-
Power supply in +5V Bodyfat
E2PROM BASE
CODE 30
Error data in E2PROM BASE
-
Error in E2PROM BASE
E2PROM PRESSV3
CODE 31
Error data in E2PROM
PRESSV3
-
Error in E2PROM PRESSV3
E2PROM BASE & PRESSV3
CODE 32
Error data in E2PROM BASE
and PRESSV3
-
Error in E2PROM BASE Y PRESSV3
ERROR PRINTER
CODE 40
Failure in PRN
-
Comms error in PRN
-
Power failure in PRN