Page 1: ...S SB B S Se er ri ie es s D Di ig gi it ta al l B Be en nc ch h S Sc ca al le es s Instruction Manual Revision 1 3 April 5 2000 SB 30 SB 70 SB 150 ...

Page 2: ...d 150 lb 70 kg capaci ties It comes standard with a bright LED screen for easy readout a removable stain less steel platform and an RS 232C serial communication port These scales can be configured to display their divisions in two separate modes non NTEP mode and NTEP mode The scale is shipped from the factory configured in NTEP mode The NTEP mode is supplied for scales which need to comply with H...

Page 3: ...he rocker switch to the ON 1 position The scale is now ready for use ON OFF SWITCH DC POWER JACK RS 232 PORT REMOTE DISPLAY ID TAG CALIBRATION SWITCH COVER SB 30 70 ONLY CONNECTOR SB 30 70 150 Back Panel The female 9 Pin RS 232 connector is used to interface to a PC Shown at right are the pin assignments for the connector Pin No Pin Name 2 Receive Data 3 Transmit Data 5 Signal Ground 2 3 5 9 pin R...

Page 4: ...enever the displayed weight is within 0 25 divisions of true zero NET Indicates that the scale is displaying net weight GROSS Indicates that the scale is displaying gross weight TARE Indicates that the tare weight in the system was established by means of the push button method lb and kg Indicates the unit of the displayed weight STABLE This light indicates that the scale is in equilibrium ...

Page 5: ...on NET GROSS Toggles between Gross and Net weight display if a tare has been established TARE Used to establish the weight on the platform as the tare weight You cannot establish zero or negative gross weight as a tare Also the scale must not be in motion or in overload lb kg Toggles between lb and kg units if enabled in the User A Menu PRINT Sends Print data to serial port if scale is not in moti...

Page 6: ...ower off to the unit 2 While holding down the lb kg key turn the power back on 3 When the display shows A1 the unit is in User Menu mode and you can release the lb kg key Shown at right are the direc tional and SET key assignments SET USER MENU CHART A1 Baud Rate 1200 2400 4800 9600 A2 Data Bits Parity 8n 7O 7E A3 Transmission Mode C d A4 Serial Data Format A5 Enable lb kg Key 0 1 0 1 2 3 4 5 7n 6...

Page 7: ...erial Data Format Selects the data format to be transmitted via the serial port to a printer or computer 0 Consolidated Controls Format 1 Toledo 8213 Format 2 A D NCI CAS Format 3 Transcell Technology Format 4 Detecto ASD Format 5 Triner Format 6 Fairbanks 70 2453 4 Format 0 1 2 3 4 5 6 A5 Disable the lb kg Key Allows the lb kg key to be disabled so that an operator cannot acci dentally press the ...

Page 8: ...ctional and SET key assign ments SET SETUP MENU CHART F1 Display Mode F6 Zero Range F9 Calib Unit 0 1 F11 Display Check F12 Calibration Press ZERO key to begin Press ZERO key to begin 0 1 100 1 9 To place the unit back into the Normal Operating mode 1 With the scale on or off toggle the Calibration Switch back to its original position Turn the scale back on without pressing any keys 2 The display ...

Page 9: ...ration Places scale into the calibration routine Scrolling down with the ZERO key one level begins the procedure Press ZERO key to begin sequence F13 Factory Reset This sub menu will reset all parameters in the F and A menus to the default settings USE WITH CAUTION Press the ZERO key twice to exe cute Calibration ON OFF SWITCH DC POWER JACK RS 232 PORT REMOTE DISPLAY ID TAG CALIBRATION SWITCH COVE...

Page 10: ...nged by pressing the PRINT key or the TARE key 5 After setting the exact value press the NET GROSS key to save the value 6 If the calibration was successful the display will show ECAL momentarily then freeze Exit the Calibration mode and enter the Normal Operating Mode by positioning the Calibration Switch back to its original position 7 If the calibration was not successful one of the error messa...

Page 11: ... x 0 05 lb 0 70 x 0 02 kg CONSTRUCTION Base Steel and ABS Platform Stainless Steel Feet Non skid Hard Rubber DISPLAY 6 Digit 0 6 7 Segment Green LED OVER CAPACITY ANNUNCIATION 102 of Full Scale Capacity OPERATING TEMPERATURE RANGE 32 F to 104 F 0 C to 40 C PLATTER SIZE SB 30 70 10 x 13 SB 150 13 x 13 POWER SOURCE AC Adapter 12VDC 500mA WEIGHT Net Weight 24 3 lb 11 kg Shipping Weight 27 8 lb 12 6 k...

Page 12: ...elect the proper settings for the following using the scale s settings Baud Rate Data Bits Parity Stop Bits 1 4 Enable the CAPS LOCK key on your PC 5 This test verifies that the scale is transmitting properly If the scale is set to Demand mode pressing the PRINT key on the scale should echo the weight readout on the PC screen If the Continuous mode has been selected on the scale the PC screen will...

Page 13: ...sary Try re calibrating the scale if this doesn t solve the problem Otherwise possible load cell damage due to overloading LO Warmup Mode The scale has been turned on without the platter installed Install the platter and then press the ZERO key if neces sary Try re calibrating the scale if this doesn t solve the problem Otherwise possible load cell damage due to shock loading Err 0 Calibration Mod...

Page 14: ...such goods by Seller discloses that the nonconformity actually exists and was not caused by accident misuse neglect alteration improper installation improper or un authorized repair or improper testing 5 Seller will have a reasonable time to repair or replace such goods THESE WARRANTIES EXCLUDE ALL OTHER WARRANTIES EXPRESS OR IMPLIED ORAL OR WRIT TEN INCLUDING WITHOUT LIMITATION WARRANTIES OF MERC...