background image

DATUM ELECTRONICS

   M425 TORQUE TRANSDUCER HANDBOOK

17

M425 mass and dimensions (mm)

A

B

C

D

E

F

G

H

I

J

M425 

model 

size

Total 

mass 

Kgs

Rotor 

mass 

Kgs

Overall 

length

Body 

length

Body 

dia.

Output 

module 

length

Output 

module 

height

Exposed 

shaft 

length

Shaft  

dia. 

G6 fit

Keyway 

length

Keyway 

width

Keyway 

depth 

off centre

Size 1 - A

0.83

0.246

184

130

66

112

3.5

27

15

22.5

5

4.3

Size 1 - B

0.83

0.253

184

130

66

112

3.5

27

15

22.5

5

4.3

Size 1 - C

0.83

0.269

184

130

66

112

3.5

27

15

22.5

5

4.3

Size 1 - D

0.83

0.291

184

130

66

112

3.5

27

15

22.5

5

4.3

Size 2 - A

2.09

1.194

240

150

85

112

3.5

45

30

44

8

11

Size 2 - B

2.09

1.288

240

150

85

112

3.5

45

30

44

8

11

Size 3 - A

6.90

4.097

315

150

110

112

5.5

82.5

50

78.5

12

20

Size 3 - B

6.90

4.439

315

150

110

112

5.5

82.5

50

78.5

12

20

Size 4 - A

20.67 12.508 425

200

135

112

5.5

112.5

75

78.5

20

30

Size 4 - B

20.67 13.807 425

200

135

112

5.5

112.5

75

78.5

20

30

Size 5 - A

38.48 26.889 416

173.7

170

112

5.5

121

110

116

32

44

Size 5 - B

38.48 27.699 416

173.7

170

112

5.5

121

110

116

32

44

Size 5 - C

38.48 28.585 416

173.7

170

112

5.5

121

110

116

32

44

Size 5 - D

38.48 29.422 416

173.7

170

112

5.5

121

110

116

32

44

A

B

C

D

E

F

K

L

M

N & O

M425 

model 

size

Total 

mass 

Kgs

Rotor 

mass 

Kgs

Overall 

length

Body 

length

Body 

dia.

Output 

module 

length

Output 

module 

height

Exposed 

shaft 

length

Spline  

outer 

dia.

Spline 

inner 

dia.

Spline 

width

Chamfer 

and 

radius

Size 6 - A

44.93 31.842 416

155

228.6 112

5.5

101.6

152.4

137

38

N = 0.5 - 1,      O = 1.1 - 1.3

Size 6 - B

44.93 34.669 416

155

228.6 112

5.5

101.6

152.4

137

38

N = 0.5 - 1,      O = 1.1 - 1.3

Table No. 9

SPECIFICATIONS 

continued

B

D

F

F

E

A

Top

Side

End

Bottom

I

B

C

H

J

H

Diagram No. 17

3D models and STEP files are  
available from Datum Electronics  
to assist project planning.  
Please contact Datum Electronics  
for more information.

This drawing and its associated design is the property of Datum Electronics Ltd. and may not be copied or used for any purpose other than that for which it is supplied, without the express written authority from Datum Electronics Ltd. 

Alphabet key for  

M425 model sizes 1 to 5

K

M

N

O

L

Letter designation key for  

M425 model size 6

Diagram No. 18

Summary of Contents for M425

Page 1: ...CTRONICS M425 TORQUE TRANSDUCER HANDBOOK AND INSTALLATIONGUIDE DATUM ELECTRONICS LIMITED TELEPHONE 44 0 1983 28 28 34 FAX 44 0 1983 28 28 35 EMAIL support datum electronics co uk WEB www datum electronics co uk ...

Page 2: ...ER INTERFACES 9 Universal interface 9 Graphic user interface 9 Versatile bulkhead mounted indicator 10 Alternate connection options 10 DATA OUTPUT 11 The M425 Transducer 3 operational modes 11 Data output format 11 Raw data torque calculation 11 Software commands direct to the M425 stator 11 How checksum is calculated 12 Sample rates and resolution 12 PC port limitations 12 OUTPUT PERFORMANCE 13 M...

Page 3: ...he M425 also has a legacy mode so that it can be used as a direct replacement for the previous M420 Transducer CHECK LIST Included as standard with the M425 Torque Transducer DESCRIPTION QUANTITY 1 M425 Torque Transducer 1 2 M425 handbook 1 3 Quick start guide 1 4 Test certificate 1 5 Datum software 1 6 Transducer cable 1 7 Power supply 1 Optional extras available from Datum Electronics DESCRIPTIO...

Page 4: ... The M425 supports a variety of universal interface models Once you are familiar with the transducer and its outputs continue to install as normal If any questions arise at this stage please call our product support team for advice The Datum Universal Interface display The Datum Universal Interface has a built in display which you can use for familiarisation with a direct connection to the M425 To...

Page 5: ...jected to the proof load level within its testing cycle Loading the Transducer above the proof level will offset the Transducer zero and will damage the Transducer Regular loading beyond the proof level will start to show progressive zero movement and may effect both the gain and the hysteresis of the Transducer The shaft absolute load is the maximum before the shaft will yield You should avoid an...

Page 6: ...as using the data cable may damage the connection The data cable is not designed to take a load Cable tie or lanyard WHICH WAY ROUND TO MOUNT The M425 Transducer will operate in both a clockwise or anti clockwise direction The M425 Transducer is calibrated to give a positive output for clockwise torque and a negative output for counter clockwise torque The M425 Transducer will also output torque d...

Page 7: ... data cable from the M425 Transducer to an interface must meet the following specification 4 core wire braided screened 7 02mm PVC Sheathed cable or equivalent The M425 Transducer current consumption is less than 250mA with a 12Vdc supply The M425 Transducer complete with its Universal Signal Interface will consume below 450mA Note alignment notch at top of contact carrier Max Cable length 200m Pi...

Page 8: ...ial torque sensor D connector Alternative power input terminal Serial Torque Sensor Factory Reset Zero Reset Ethernet USB Type A Data Logging USB Mini B Primary USB to PC InterfaceMain Power Input Serial Output Diagram No 13 Universal Interface connections guide Power Analogue 2 way Speed Analogue 2 way Torque Analogue 2 way Thrust Analogue 2 way 12 Volts OUT 2 way Serial IN copy 2 way Tacho Pulse...

Page 9: ...ill also accept an input from an external tachometer sensor the output from this sensor can be directly linked into the torque and power data USER INTERFACES The outputs that can be configured are Datum data logging software The Datum data logging and configuration software provides a wide range of control functions of the M425 including set up and configuration options plus data logging GRAPHIC U...

Page 10: ...ndard ALTERNATE CONNECTION OPTIONS The M425 is compatible with a variety of other universal interfaces or indicators by correct cable and PIN configuration This allows direct connection to their user software and customers own software See page 7 for a guide to wiring connector configuration The M425 will also accept configuration commands to enable connection to legacy Datum Electronics interface...

Page 11: ...sensor 25 6 is the RPM CS is the data checksum This type of string containing multiple mV V signal value is used where the baud rate of the receiving device is limited and a higher data rate is required String type B ZR 0 0492 0 0492 0 0492 0 0492 0 0492 0 0492 0 0492 0 0492 0 0492 0 0492 24 1 21 6 CS Where ZR is the string identifier 0 0492 are the last 10 signal values in mV V from the sensor 25...

Page 12: ...in tables 1 and 2 on page 13 These table show the rates for a range of standard output strings PC PORT LIMITATIONS Note the M425 Transducer is capable of transmitting at the rates above Some PC s have limited port configurations that may not be able to cater for data rates at this level The maximum data rate at a given baud rate can be calculated as The basic string with a single torque value and ...

Page 13: ...00 38400 57600 115200 256000 460800 921600 3000000 Number of torque values Torque readings per second Example of the output string Type 1 1 per packet 43 87 174 260 525 1100 2000 4000 4000 ZR T R C V CS Type 2 5 per packet 88 177 355 533 1066 2000 4000 4000 4000 ZR T T T T T R C V CS Type 3 10 per packet 102 204 408 612 1225 2000 4000 4000 4000 ZR T T T T T T T T T T R C V CS Type 4 16 per packet ...

Page 14: ... 0778 0 0779 0 0781 0 0781 0 0783 0 0781 0 0781 5 8E 08 221 2b ZF 0 0781 0 078 0 0785 0 0784 0 0785 0 0782 0 0779 0 078 0 0781 0 0779 0 0779 0 0778 0 0776 0 0779 0 0778 0 0775 5 8E 08 222 37 ZF 0 0781 0 0771 0 0776 0 077 0 0768 0 0768 0 0772 0 0772 0 0769 0 0766 0 0769 0 0766 0 0763 0 077 0 0767 0 0764 5 9E 08 223 15 ZF 0 0762 0 0759 0 0764 0 0761 0 0763 0 0761 0 0762 0 0761 0 076 0 0763 0 076 0 0...

Page 15: ...S425 for IP67 68 Electromagnetic Compatibility EN61326 1 2006 IEC61000 4 IEC60945 Table No 7 SIGNAL INTERFACE OUTPUTS Digital RS485 Serial data RS232 Serial data Ethernet Data and configuration MODBUS UDP Serial data USB Type Mini B Serial data USB Type A Data logging Analogues All four of the M425 Transducer s analogue output channels can be configured for any of the following settings by the use...

Page 16: ...A 5000 3687 9 0 3500 425 218 75 2364 7477 13353 24630 2 22E 06 450056 Size 4 B 10000 7375 7 0 3500 425 218 75 4618 14604 26079 38485 1 27E 06 789591 Size 5 A 15000 11064 0 2000 416 191 8502 23477 41922 52810 3 82E 07 2614430 Size 5 B 20000 14751 0 2000 416 191 11241 31040 55428 63617 2 95E 07 3390520 Size 5 C 25000 18439 0 2000 416 191 14513 40075 71562 75430 2 39E 07 4185528 Size 5 D 30000 22127 ...

Page 17: ...5 B 38 48 27 699 416 173 7 170 112 5 5 121 110 116 32 44 Size 5 C 38 48 28 585 416 173 7 170 112 5 5 121 110 116 32 44 Size 5 D 38 48 29 422 416 173 7 170 112 5 5 121 110 116 32 44 A B C D E F K L M N O M425 model size Total mass Kgs Rotor mass Kgs Overall length Body length Body dia Output module length Output module height Exposed shaft length Spline outer dia Spline inner dia Spline width Chamf...

Page 18: ...lied with an indicator or a signal interface the outputs of the interface and the Transducer will both be calibrated and detailed on a systems test certificate The outputs calibrated will be the 10Vdc 5Vdc and the 4 20mA signals The indicators and serial data outputs for the Ethernet USB and MODBUS ports are directly sent from the Transducer Interface options allow a scaling factor s for these sig...

Page 19: ...or 10 seconds will take the current torque reading as zero When checking or setting the zero check to see if any torque is locked into the transducer by the machine or rig on which it is used It is ideal to check the zero with the transducer disconnected from the rig NO DATA If you have no data being transmitted from the Transducer check the LED on top of the Transducer at the opposite end to the ...

Page 20: ...ords transducer and sensor are often used in this context to mean the same thing Span This is the value of output at the rated load Either given in terms of mV V signal or Nm or lbft when in engineering units STEP files A STEP file is a widely adopted CAD file format used to share 3D models between users with different CAD systems Torque The twisting force on the shaft created by the driving force...

Page 21: ...M425 TORQUE TRANSDUCER HANDBOOK 21 DATUM ELECTRONICS LIMITED TELEPHONE 44 0 1983 28 28 34 FAX 44 0 1983 28 28 35 EMAIL support datum electronics co uk WEB www datum electronics co uk DESIGNED BY WWW PEPPERCREATIVE CO UK ...

Reviews: