S/5 Hemodynamic modules
Document No. 800 1008-2
57
3.
Connect an external mercury manometer with pump to module through the both tubes of the
hose. Pump up to about 200 mmHg pressure (range of 150 to 300 mmHg allowed) according
to the manometer. Verify that both pressure values in the prompt field match the manometer
reading. If not, adjust by turning the ComWheel.
4.
When the values are equal, push the ComWheel to confirm the calibration. First the message
‘Calibrating’ will appear in the digit fields for NIBP followed after a few seconds ‘Calibrated’,
which means that the calibration data has now been saved.
5.
Use the bottom switch to enable
Protection
setting and set it ON, and finally disable
Protection
setting.
5.1.3
NIBP safety valve menu
Start Test
Start test
is for starting and
Stop test
is for stopping the Safety Valve test.
Safety Valve Data
See NIBP Service menu in chapter
for information on general items
Pressure, Zero, Protect
handle, Calibr. prot., +15 V power, AD0
to
AD7
as well as
Timeouts
etc.
Max. press
and
2 s after stop
show the measured values at Safety Valve test.
Safety Valve Test Adult/Infant
Wrap an adult cuff around a pipe and connect the cuff to the module. Highlight
Start test
and give
the ComWheel a push. The test ends automatically or when
Stop test
(appears in place of
Start
test
) is pushed.
Max. press
indicates the pressure at which the safety valve opens and is normally 310
±
15 mmHg
for adult and 150 mmHg
±
15 mmHg for infant.
2 s after stop
indicates the pressure at 2 seconds
after the pump has stopped and is normally > 280 mmHg for adult and > 120 mmHg for infant. If
the value is less, check leakage by the active leak test.