Datex-Ohmeda S/5 monitors
Document No. 800 1008-2
44
TROUBLE
CAUSE
TREATMENT
Cuff loose-message
1. Hose and/or cuff not connected.
1. Connect the hose and the cuff.
2. Hose and cuff connected.
Reasons:
2.
−
cuff loosely wrapped
−
tighten the cuff
−
leakage in cuff or hose
−
replace cuff/hose
−
leakage inside module
−
check internal tubing and air
chamber, and fix if necessary
−
pump does not work
−
check pump connector; if OK, replace
pump
−
no pulses during the last three
measurements
−
check cuff positioning
Air leakage-message
1. Hose or cuff leaking. Reasons:
1. Replace cuff
−
cuff damaged
−
replace cuff
−
cuff connector damaged
−
replace cuff connector (if the fault is
in hose connector,)
−
O-ring damaged or missing
−
replace O-ring
−
hose double connector
damaged
−
replace hose
2. Hose and cuff OK. Reasons:
2. Connect or replace tube
−
leakage inside the module
−
replace the whole tubing
−
tube disconnected or damaged
−
fix connections
−
air chamber leaking
−
−
tubes or valve(s) damaged
−
replace tubes/valve(s)
Unable to measure Sys-
message
Systolic blood pressure probably
higher than the maximum inflation
pressure.
Automatic retrial with increased
pressure.