![Datex-Ohmeda S/5 M-ESTPR Technical Reference Manual Slot Download Page 17](http://html.mh-extra.com/html/datex-ohmeda/s-5-m-estpr/s-5-m-estpr_technical-reference-manual-slot_513843017.webp)
S/5 Hemodynamic modules
Document No. 800 1008-2
11
boards, attached to the front panel of the module. The front panel has six connectors and four keys.
The connectors are two for temperature measurement, two for invasive blood pressure
measurement, one for ECG, and one for SpO
2
measurement. The keys are for ECG lead, Start
Wedge, P1 zero, and P2 zero.
In
M-ETPR module
,
there are two small boards, the ECG input board and the 2P input board
attached to the front panel of the module. The front panel has five connectors and four keys. The
connectors are two for temperature measurement, two for invasive blood pressure measurement,
and one for ECG
measurement. The keys are for ECG lead, Start Wedge, P1 zero, and P2 zero.
In
M-ESTR module
, there are two small boards: the S input board and the ECG input board,
attached to the front panel of the module. The front panel has four connectors and one key. The
connectors are two for temperature measurement, one for ECG, and one for SpO
2
measurement.
The key is for ECG lead select.
NOTE: M-ESTP rev. 03, M-ETP rev. 02 and M-EST rev. 02 and all earlier revisions have separate T
and SP input boards.
2.2.2
M-NE12STPR/-NE12STR/-NE12TPR/-NESTPR/-NESTR/-NETPR modules
Figure 4
Front panel of M-NESTPR
The
M-NESTPR, M-NETPR, and M-NESTR
modules
contain three main PC boards, the STP board,
the ECG board, and the NIBP board. They work independently. Each of these has their own
processor and software EPROM.
The
M-NE12STPR, M-NE12TPR, and M-NE12STR
contain three main PC boards, The STP board,
the ECG board and the NIBP board. They work independently. Each of them has their own
processor. The STP board and NIBP board have software EPROM. In the ECG board the software is
in flash memory. The STP and NIBP boards are the same as in M-NESTPR module but the ECG
board and ECG input board are different.
In
M-NESTPR module
, there are two small boards, the SP input and the ECG input board attached
to the front panel of the module. The front panel has seven connectors and four keys. The
connectors are two for temperature measurement, two for invasive blood pressure measurement,
one for ECG, one for NIBP, and one for SpO
2
measurement. The keys are for NIBP Auto On/Off,
NIBP Start/Cancel, P1 zero, and P2 zero. The structure of
M-NE12STPR
is similar except the ECG
board and ECG input board are different.
T2
P1
P2
T1
NIBP
ECG+
Resp
Auto
On/Off
Zero
P1
Zero
P2
Start
Cancel
SpO
2
Summary of Contents for S/5 M-ESTPR
Page 2: ......
Page 6: ...Datex Ohmeda S 5 monitors Document No 800 1008 2 iv This page intentionally left blank ...
Page 83: ...APPENDIX A Service check form S 5 Hemodynamic modules Document No 800 1008 2 77 APPENDIX A ...
Page 84: ...Datex Ohmeda S 5 monitors Document No 800 1008 2 78 This page intentionally left blank ...