Datex-Ohmeda S/5 monitors
Document No. 800 1008-2
36
29.
Check the NIBP tubing system for leakages.
Select
Calibrations
from the NIBP service menu.
Connect the pressure manometer to the NIBP module cuff connector. Start the active leak
test from the menu by pressing the ComWheel. The module pumps a pressure of about 265
mmHg and then the pump stops.
Wait for 15 seconds for the pressure to stabilize then check that the pressure does not drop
more than 5 mmHg per one minute. Release the pressure by pressing the ComWheel once
more.
?
30.
Calibration check.
Disconnect the pressure manometer. Select
Calibrations
and then highlight
Calibration
Check
. Press the ComWheel and take down the zero offset values for both pressure
transducers, B1 and B2. The values should be within ±10 mmHg.
Connect the pressure manometer to the cuff connector and check the calibration with
pressures 100 mmHg, 200 mmHg and 260 mmHg. The zero offset value must be added to
the displayed pressure value in order to determine the real pressure.
Recalibrate the NIBP measurement according to the instructions in the chapter 3.4
Adjustment and calibration,
, if necessary. Remember to set the calibration
protection back on after the calibration.
?
31.
Check the watchdog timer activation pressure.
Select
Pneumatics
from the NIBP service menu.
Keep the pressure manometer connected to the cuff connector. Pump up the pressure very
slowly and note the value on the manometer when your hear a signal from the loudspeaker.
The pressure at where the watchdog timer should activate with an audible signal is
7.5 mmHg (5...10 mmHg). Adjust the limit with the trimmer on the NIBP board, if necessary.
?