DT4120 xx xx (PS)
88
20170324-V0
10.3. Setting up the instrument (example)
Remarks
It
is
an
additional example for
Chapter 7. Setting-up.
It
helps to understand the setting of values through a practical
example. The steps of the whole setting you will find in
Chapter 7.
.
Exercise
Setting parameters
The task is: temeperature measurement with Pt100 sensor
with 4 wire connection.
A DT4120 xx xx (PS) has an analog output option, output
level 4-20 mA, scaleing: 0 °C ÷ 4 mA, 100 °C ÷ 20 mA.
Limit 1 is active when T > 90 °C, hysteresis: ±0.1 °C.
Leading zeroe’s are disabled.
The DT4120 xx xx (PS) parameter settings:
Here are listed only those steps what are differ from Factory
default:
•
06.NUL =
(leader zeros are not displayed)
•
10. L1: L1.SET =
(limit output is active when
displayed value > limit value)
•
10.L1: L1.L:LI =
(limit value 90.0 °C)
•
10. L1: L1.HYS =
(hysteresis ±0.1 °C)
•
15. A.S.L =
(analog output 4 mA ÷ 0.0 °C)
•
16. A.S.H =
(analog output 20 mA ÷ 100.0 °C)
Summary of Contents for DT4120 Series
Page 1: ...DT4120 xx xx PS Temperature Resistance Potentiometer Meter Transmitter Operating Instructions...
Page 84: ...DT4120 xx xx PS 84 20170324 V0 10 2 Application example Analog output option...
Page 85: ...DT4120 xx xx PS 20170324 V0 85 RS232 communication interface option...
Page 86: ...DT4120 xx xx PS 86 20170324 V0 RS485 communication interface option...
Page 87: ...DT4120 xx xx PS 20170324 V0 87 RS485 bus topology...
Page 90: ...DT4120 xx xx PS 90 20170324 V0...
Page 91: ...DT4120 xx xx PS 20170324 V0 91...
Page 92: ......