7.16. Assignment analog output low (start) value to
display value
(scaling)
(15. menu item)
Function
Here
you
can
assign
analog output start value
(0 mA or 4 mA) to the measured (temperature, resistance,
potentiometer position) value. This way you can assign any
phisical range to any output range.
[Factory default: -200.0 °C. so in case of a measured value
-200.0 °C the output value will be 4 mA (or 0 mA).]
Notice!
For scaling the analog output must be assign the
analog output high (end) value also. (Next menu item)
Sequence of operations
1. Enter the menu by typing the supervisor code.
The way the code should be type in is described in Chapter
7.1. Typing the code in.
You see this on the display:
.
2. Keep stepping by pressing the
▲
button as long as you
see this menu item:
.(
A
nalog
S
cale
L
ow = analog
scale low value).
3. Enter the menu item by pressing the
ENTER
button.
4. The number:
. is shown on the display. The
number is blinking, signaling that it can be modify.
5. Pressing the
►
button you can select the next digit.
6. Pressing the
▲
button you can increase the value of the
blinking digit:
,
,
,
,
,
,
,
,
,
,
, etc.
7. When the digit on the left side is selected pressing the
▲
button the sequence is as follows:
,
,
,
,
,
,
,
,
,
,
,
, etc. so you
can set any number between -9999-99999.
Summary of Contents for DT4120 Series
Page 1: ...DT4120 xx xx PS Temperature Resistance Potentiometer Meter Transmitter Operating Instructions...
Page 84: ...DT4120 xx xx PS 84 20170324 V0 10 2 Application example Analog output option...
Page 85: ...DT4120 xx xx PS 20170324 V0 85 RS232 communication interface option...
Page 86: ...DT4120 xx xx PS 86 20170324 V0 RS485 communication interface option...
Page 87: ...DT4120 xx xx PS 20170324 V0 87 RS485 bus topology...
Page 90: ...DT4120 xx xx PS 90 20170324 V0...
Page 91: ...DT4120 xx xx PS 20170324 V0 91...
Page 92: ......