37
2.
Establish communication with the “Probe”
Because the CMS software does not establish connection with the color calibration device
automatically, you will need to manually establish communication between the Probe and the PC.
Select “Tools
Calibration
Calibration Interface or Display Characterization”
The CMS should detect the probe and a dialogue box with the message “Do you want to connect
to i1 Display Pro”should pop up.
Note: If the probe model is CA-310, you should see the corresponding model number in the
message.
Click “OK” if the detected color calibration device is correct.
3.
Warm up the monitor before calibration.
Make sure the monitor and the color calibration device are turned ON and leave them running for
about 30 minutes.
4.
Enable the TLM-170K monitor’s color calibration mode.
Open the TLM-170K’s OSD menu, then select System
Color Calibration
ON.