Calibration Verification (U.S. Pounds)
Product Reference Guide
5-7
Calibration Verification (U.S. Pounds)
Once you have completed the calibration sequence, you may be required
to perform these step-by-step verification procedures. These procedures
follow the
National Institute of Standards and Technology Handbook-44
guidelines for grocery scale installations. You may be required by state or
local law to have these procedures performed by a certified technician or
verified by a proper official.
These calibration verification procedures cover five different tests:
•
Increasing-Load Test
•
Shift Test
•
Blanking Test
•
Decreasing-Load Test
•
Return to Zero Test
Increasing-Load Test (Phase 1)
This test checks Scale operation with five, ten and fifteen pound weights.
1. Check that the display reads 0.00 lb. when at rest with nothing on
the weighing surface. (The Yellow LED is steadily lit).
2. Place a five-pound weight on the center of the weighing surface and
check that the display reads 5.00 lb.
3. Place an additional five pounds on the center of the weighing sur-
face and check that the display reads between 9.99 and 10.01 lb.
4. Place an additional five pounds on the center of the weighing sur-
face and check that the display reads between 14.99 and 15.01.
Remove the weights and verify that the display reads 0.00.
5. You have completed the Increasing-Load Test (Phase 1).
Summary of Contents for Magellan 9500?
Page 1: ...Magellan 9500 9500 9504 9504 with SmartSentry Product Reference Guide...
Page 28: ...1 16 Magellan 9500 NOTES...
Page 100: ...4 14 Magellan 9500 NOTES...
Page 116: ...5 16 Magellan 9500 NOTES...
Page 396: ...6 280 Magellan 9500 NOTES...
Page 416: ...B 10 Magellan 9500 NOTES...
Page 426: ...D 2 Magellan 9500 NOTES...
Page 442: ...E 16 Magellan 9500 NOTES...
Page 458: ...F 16 Magellan 9500 NOTES...