background image

Copyright © 2019 

 2020 DARE!! Products 

Page 20 

 

 

 

 

can be: "ON" or "OFF". 

MEAS?<space>LSRTEMP 

Get the measured laser 
temperature in degrees 
Celsius. 

<temperature>  
For example: "35.6". 

MEAS?<space>LSRCURR 

Get the measured laser 
current in Amperes. 

<laser current>  
For example: "0.051". 

MEAS?<space>INT 

Get the measured voltage 
of the internal supply in 
volt. 

<supply voltage>  
For example: "14.400". 

MEAS?<space>EXT 

Get the measured voltage 
of the external supply in 
volt. 

<supply voltage>  
For example: "12.000". 

SUPPLY<space> INT 

Switches the remote unit to 
internal battery. 

OK or Error code. 

SUPPLY<space> EXT 

Switches the remote unit to 
external power supply. 

OK or Error code. 

RF<ch> 

Set the RF switch to the 
<ch> channel.  
<ch> can be 1 

 8. 

Note: channel is active 
350ms after this command 
is acknowledged by the 
RadiLink. 

OK or error. 

RF? 

Returns which channel is 
selected. 

For example: 3. 

GAIN<space><gain> 

Set the gain of the system. 
<gain> can be STD or HIGH. 

OK or error. 

GAIN? 

Returns the gain setting. 

<gain> can be: 
STD for typical 0 dB gain. 
HIGH for typical 20 dB gain. 

INFO? 

Returns additional error 
states. 

<error>  
For example: "ERROR 35". 

 

Command 

Description 

Reply 

BIAS<ch><space> 
STATE,<state> 

Turn the supply of a 
channel on or off. 
<state> can be ON or OFF. 

OK or error. 

BIAS<ch>? <space> 
STATE 

Returns the state of the 
bias channel.  
<state> can be ON or OFF 
or error code that the 
current limit has occurred.  

ON or OFF or error. 

BIAS<ch><space> 
SOURCE,<source> 

Select the bias source. 
<ch> can be 1 

 8. 

OK or error. 

Summary of Contents for RadiLink RLK2008A

Page 1: ...2 Fax 31 348 430 645 www dare eu instruments dare nl Dijkstra Advice Research EMC Instruments B V KvK Utrecht 30138673 Rabobank Utrechtse Waarden 15 83 13 585 IBAN NL31RABO0158313585 SWIFT code RABONL...

Page 2: ...roduct Characteristics 9 2 2 Components 9 3 Installation 11 3 1 Connections and indicators 11 3 2 Switching ON and OFF 14 3 3 Activating the link 14 3 4 Internal battery 14 3 5 Firmware updates 16 4 S...

Page 3: ...t maintenance or repair to the equipment Verify that the supply voltage is within the operating range of the equipment The RadiLink contains materials which can be recycled and reused At the end of it...

Page 4: ...a class 1M LASER product according to EN60825 1 2014 and EN60825 2 2004 It is mandatory to check the fibre connections before turning on the system According to EN60825 2 the RadiLink is equipped wit...

Page 5: ...cause deterioration or permanent damage Please follow the handling and maintenance guidelines below Some important guidelines for handling of the fibres and connectors are Always place the plastic end...

Page 6: ...not use other solvents or wipes Put some alcohol on the wipe and rub gently over the contact surface of the fibre connector Dry the surface with a dry part of the wipe Use the examples and instruction...

Page 7: ...imum Furthermore common mode currents flowing through shields of coaxial cables are eliminated by the use of this fibre system This 8 channel RadiLink system is a special version designed for automoti...

Page 8: ...d computer interface for all the RadiCentre plug in cards and modules The RadiCentre is available as multislot system with touch screen Or as a single slot system without display RadiMation software R...

Page 9: ...to switch between different inputs without the need to alter the test setup between measurements Support for active antennas All RF inputs can provide a bias supply for active antennas This supply can...

Page 10: ...or connecting the remote unit to the plug in card Product code CBL2003 25M Lithium Ion charger plug for charging the internal battery of the remote unit With 3 way Lemo plug FGG 1B 303 CLAD62Z Supply...

Page 11: ...ators Green LED for Power On SMA connector for RF output signal to measurement device Red LED for Link Active indication Optical DIN connector for Analogue Link Optical ST connector for Digital Link C...

Page 12: ...or charging the internal Lithium battery use the supplied charger with the 3 way Lemo plug to charge the internal battery For powering the remote unit from an external source use the supplied power ca...

Page 13: ...C voltage is present on the RF input connector for the active antenna 4x channel active indicators Green LED o these LED s will illuminate to indicate the channel which RF signal is transferred over t...

Page 14: ...for the Batt low and Link Act LED to blink at the same time and release the button The remote unit will now switch OFF 3 3 Activating the link Once the remote unit is powered on the LASER link can be...

Page 15: ...Discharging Lithium cells must never be discharged below a certain voltage to avoid the possibility of irreversible damage Therefore the remote unit is equipped with a circuit that shuts down the remo...

Page 16: ...tion lite This can be done from the advanced device driver settings selecting the Software update tab The RadiLink system must be completely switched on prior to starting the firmware update At the en...

Page 17: ...Equivalent input noise 150kHz 30MHz 30MHz 6GHz Typical 18 dB V 9kHz RBW Typical 8 dB V 120kHz RBW Input VSWR 1 2 max 150kHz 1 GHz 1 4 max 1 6 GHz Output VSWR 1 4 max 4 2 Bias Internal bias voltage 5 0...

Page 18: ...cells n a Lithium charger Ansmann 9C94212 n a 4 7 Optical Digital LASER Plug in card Max 2 mW 1310 nm Digital LASER remote unit Max 2 mW 1550 nm Analogue laser remote unit Typical 2mW max 5 mW 1310 n...

Page 19: ...LEAR Clears all present errors OK The following command can be used to control the remote unit or obtain data from it Command Description Reply STATUS Returns the status of the system The reply can be...

Page 20: ...OK or Error code RF ch Set the RF switch to the ch channel ch can be 1 8 Note channel is active 350ms after this command is acknowledged by the RadiLink OK or error RF Returns which channel is select...

Page 21: ...BIAS ch space LVL Returns the level of the internal bias source For example 6 000 BIAS ch LVL space MIN Returns the minimum level of the internal bias source For example 5 000 BIAS ch LVL space MAX R...

Page 22: ...T for continues mode and PULSE for pulse mode RELAY index space PULSE value Set the pulse duration index can be 1 or 2 value is time in seconds that the pulse is active Example 0 5 for half a second O...

Page 23: ...too high temperature 165 RLR Laser off through too long no communication 166 RLR Amp adjustment fail 167 PML Laser intensity adjustment fail 168 PML Laser modulation adjustment fail 169 RLS not detec...

Page 24: ...thin the warranty term The customer will need to cover the costs for returning the product to DARE such as shipping and or any applicable duties and taxes DARE Instruments will arrange the courier and...

Page 25: ...ons of the applicable requirements of the following harmonized standards Emission EN 61326 1 2013 Class A1 Immunity EN 61326 1 2013 Industrial level performance criteria A Safety EN 61010 1 2010 EN 60...