![DARAY V460S Series User Manual & Installation Manual Download Page 41](http://html1.mh-extra.com/html/daray/v460s-series/v460s-series_user-manual-and-installation-manual_3483296041.webp)
Page
40
of 49
7.5 BP SETUP
From the system menu select the [BP SETUP] key and the following menu will pop up.
Auto Measure
–
In PM mode (selected via system setup), when a user selects [ON] for auto
measurement, the Monitor will measure blood pressure according to the time selected in the
[INTERVAL] item. In ABPM mode, auto measurement and the interval key turn grey and cannot be
used.
Interval -
The interval time works in minutes and defines the time between measurements; the
time interval can be selected from the following options: 5, 10, 15, 20, 30, 45, 60, and 90
minutes.
Alarms –
the alarm function for the BP menu can be set to high and low limit’s which can be setup
here. When the pressure is higher or lower than the defined limits, the alarm will activate. The
Systolic and Diastolic records can have individual alarms each with their own high and low rates.
These can be adjusted the same way as the previous panels, navigate up and down to the
selection, select it then alter the levels with the up and down keys and select OK to confirm any
changes.
When users adjust the patient parameters by selecting [ADU], [NEO] or [PED] in
the [MEASURE MODE] the alarm ranges will change as standard.
The adjustable ranges of the high and low limits of the alarm are as follows:
SYS: 40-270mmHg
DIA: 20-215mmHg
7.6 SpO
2
SETUP
When you select PM SPO
2
SETUP in the "SYSTEM MENU". The following menu will pop up.
Pulse Sound –
For the Pulse Sound you can choose its state. When the switch turns on, plug in the
relevant oximeter attachment to measure the pulse in the monitoring mode. There is no limit for
silence and alarm pause states.
Summary of Contents for V460S Series
Page 2: ...Page 1 of 49...
Page 10: ...Page 9 of 49 2 General Operation Guide Section...
Page 15: ...Page 14 of 49 4 NIBP Operation...
Page 20: ...Page 19 of 49 5 SpO2 Operation...
Page 24: ...Page 23 of 49 6 Software Operation Initial Set up Guide Section...
Page 38: ...Page 37 of 49 7 First Time Setup and the on board menu functions...