Parameter lists
A.3 Diagnostic
78
SONO 3500 CT
Operating Instructions, 07/2019, VI.SH.P2.02
Parameter
Default
value
Value
range
Access
level
Description
217
Path 2 fault counter
e.g. 4
0 ... 65535 Read
Total number of faults.
Valid only for 2-path solutions and if alarm is
enabled
218
Path 2 fault appears
Last fault
appear date
dd-mm-
yyyy,
hh:mm:ss
Read
First time the fault appeared.
Valid only for 2-path solutions and if alarm is
enabled
219
Path 2 fault disappears
Last fault
disappear
date
dd-mm-
yyyy,
hh:mm:ss
Read
Last time the fault disappeared.
Valid only for 2-path solutions and if alarm is
enabled
220
Dual slope alarm output
enable
Yes
No, Yes
R/W
Select "Yes" to enable active alarm on alarm
output / call-up
221
Dual slope fault hours
e.g. 0 h
0 ... 65535 Read
Total hours fault active.
Valid only if alarm is enabled
222
Dual slope fault counter
e.g. 0
0 ... 65535 Read
Total number of faults.
Valid only if alarm is enabled
223
Dual slope fault appears
e.g. 0 h
0 ... 65535 Read
First time the fault appeared.
Valid only if alarm is enabled
224
Dual slope fault disappears e.g. 0
0 ... 65535 Read
Last time the fault disappeared.
Valid only if alarm is enabled
225
Parameter checksum alarm
output enable
Yes
No, Yes
R/W
Select "Yes" to enable active alarm on alarm
output / call-up
226
Parameter checksum fault
hours
e.g. 0 h
0 ... 65535 Read
Total hours fault active.
Valid only if alarm is enabled
227
Parameter checksum fault
counter
e.g. 0
0 ... 65535 Read
Total number of faults. Valid only if alarm is
enabled
228
Parameter checksum fault
appears
e.g. 0 h
0 ... 65535 Read
First time the fault appeared.
Valid only if alarm is enabled
229
Parameter checksum fault
disappears
e.g. 0
0 ... 65535 Read
Password to change parameters in the
flowmeter.
Valid only if alarm is enabled
230
Low power alarm output
enable
Yes
No, Yes
R/W
Select "Yes" to enable active alarm on alarm
output / call-up
231
Low power fault hours
e.g. 0 h
0 ... 65535 Read
Total hours fault active.
Valid only if alarm is enabled
232
Low power fault counter
e.g. 0
0 ... 65535 Read
Total number of faults.
Valid only if alarm is enabled
233
Low power fault appears
e.g. 0 h
0 ... 65535 Read
First time the fault appeared.
Valid only if alarm is enabled
234
Low power fault disappears e.g. 0
0 ... 65535 Read
Last time the fault disappeared.
Valid only if alarm is enabled
235
Flow overflow alarm output
enable
Yes
No, Yes
R/W
Select "Yes" to enable active alarm on alarm
output / call-up
236
Overflow fault hours
e.g. 0 h
0 ... 65535 Read
Total hours fault active.
Valid only if alarm is enabled