Commissioning
6.4 Commissioning via PDM
46
SONO 3500 CT
Operating Instructions, 07/2019, VI.SH.P2.02
Call up indication:
When output B is configured as "call-up", the output is activated by an alarm condition and
remains on until it is manually reset via communication interface and the PDM program (or
via manually power down of the device).
A new alarm will not activate a "call-up" function if the "call-up" function is still active from a
previous alarm.
6.4.8
Checking the operation readiness
All parameters are now set and defined according to the application.
1. Select: "Device"
→ "Download to device..." to download the parameters to the device.
Note
Before downloading the parameters, check that all listed data are loaded or changed and
in accordance with the application requirements.
2. Select "Execute even if the device TAG does not match the project data TAG." and click
"OK" to download all changes from the table to the device.
View process values
The system is now ready for normal operation.
1.
Select "View" → "Display" to see all measured process v
alues.
2. Verify that the fields show the expected values.