MAGFLO
â
S
p
e
c
if
ic
a
ti
o
n
s
14
MAG 5000 or MAG 6000 used with MAG 3100 W or MAG 1100 PFA
2.6
Meter uncertainty
V: Actual flow velocity [ft./s]
E: Meter uncertainty as a percentage of actual flow
MAG 6000 used with MAG 3100, MAG 1100 Ceramic or MAG 5100 W
V: Actual flow velocity [ft./s]
E: Meter uncertainty as a percentage of actual flow
Reference conditions
(ISO 9104 and DIN/EN 29104)
Temperature of medium
68
°
F
±
9 F
Ambient temperature
68
°
F
±
9 F
Supply voltage
Un
±
1%
Warming-up time
30 min.
Incorporation in pipe section
Inlet section 10 * Nominal pipe size (sizes up to 48"),
5 * Nominal pipe size (sizes up to 48")
Outlet section 5 * Nominal pipe size (sizes up to 48"),
3 * Nominal pipe size (sizes up to 48")
Flow conditions
Fully developed flow profile
Additions in the event of deviations from reference conditions
Current output
As pulse output
±
(0.1% of actual flow +0.05% FSO)
Effect of ambient temperature
Display/frequency/pulse output: <
±
0.003% / <
±
0.0017
°
F
Current output: <
±
0.005% / <
±
0.0028
°
F
Effect of supply voltage
< 0.005% of measuring value on 1% change
Repeatability
±
0.1% of actual flow for V
³
1.5 ft./sec.
2. Specifications