25
Figure 11: Mode 2 Timing
Frame Period
Readout
Exposure Time
Frame Period
Readout
Programmable Period
Exposure Time
Programmable Period
Programmable Period
Programmable Period
Mode 3: External Trigger with Maximum Exposure
he external trigger pulses. The falling edge of the
external trigger marks the beginning of the maximum exposure with exposure control
disabled (ECD).
Figure 12: Frame Rate is set by External Trigger Pulses.
Frame rate is set by the period of t
Frame Period
Exposure Time
Frame Period
Readout
Exposure Time
EXSYNC
Readout
Mode 4: Smart EXSYNC, External Frame Rate and Exposure Time
In this mode, EXSYNC sets both the frame period and the exposure time. The rising edge of EX
the beginning of the exposure and the falling edge initiates readout. Refer to
Table 11: Allowable Exposure Time Increments on page 27 for details on minimum
exposure time increments for this mode.
Figure 13: Trigger Period is Repetitive and Greater than Read Out Time.
SYNC marks
EXSYNC
Frame Period
Frame Period
Exposure Time
Readout
CR
Waiting
Exposure Time
Readout
CR
Waiting
CR= Charge Reset
DALSA 03-32-10121-09