System precautions
The following are precautions that must be taken in order to prevent possible damage to
your Teledyne DALSA Detector:
Warning:
The use of accessories, power supplies and cables other than those specified, with the
exception of cables sold by Teledyne DALSA as replacement parts, may result in
increased emission or decreased immunity of the Shad-o-Scan 3001/4501 Detector.
Static precautions
Observe proper ESD/static control procedures when handling system components. The
use of properly grounded wrist strap is highly recommended.
Installation Precautions
Never connect or disconnect cables while power to the detector is on. Damage to the
detector electronics might occur if the cables are connected and/or disconnected while
the detectors power is on.
Service by Qualified Personnel Only
This equipment is to be installed by a qualified technician only. The system can only be
used in conjunction with a properly installed X-ray source with the appropriate shielding
and a properly configured computer workstation that meets the minimum system
requirements discussed later in this manual.
Environmental Specifications
Temperature
-
Operating
-
Transport & Storage
+10..+40
o
C
-10..+55
o
C
Humidity (non-condensing)
-
Operating
-
Transport & Storage
10%...80%
10%...80%
Shock
25G (6ms)
Vibration
1G (10-150Hz)
IP Class
IP69K
Note: For transport and storage, the standard product packaging is assumed to be in