32
Shad-o-Scan 3001/4501 Detector
Teledyne DALSA
Triggered (External) Mode
In this mode the detector will take a single image at the moment a trigger signal is
applied to the detector.
When using pulsed X-rays, the X-ray exposure including X-ray rise- and fall-times have
to be within the “Expos
ure Time
” window
(C).
The trigger output indicates the Exposure Time window when no read-out of the sensor
occurs.
Row #
Pixel Array
Time
ReadOut
Integration
Time
Frame (N)
Frame (N+1)
Trigger Input
Frame Blanking
interval
Sensor Data Output
C
External Trigger Mode
A
Operation mode
Full resolution Binning 2x2
1/A: Frame rate = trigger frequency 730 FPS
1390 FPS
A: integration time
1.36 ms
0.644 ms
B: readout time
1.36 ms
0.644 ms
C: Frame Blanking interval (A_B)
0.001 ms
0.001 ms
Figure timing diagram 'Trigger Mode'