Shad-o-Scan User Manual, Rev04
15
Shad-o-Scan 3001/4501 X-Ray Detector
Teledyne DALSA
Power Connector
1
2
Pin 1 indicator
Rear View
Plug
Power Connector
Description
1
Supply voltage (+11-16V / 21W)
2
Ground
Shielding (chassis)
The detector requires a single voltage input (+11-16V / 21W). There is a protection for
misconnection, via a resettable fuse. When the fuse has been activated (isolating),
power cycling will return it to conductive state again thus restoring normal operation
(assuming power has been properly applied).
WARNING: When setting up the detector power supplies follow these guidelines:
•
Apply the appropriate voltages.
•
Keep leads as short as possible in order to reduce voltage drop. The voltage
level at the product power connector should be within the range mentioned
in the datasheet.
•
Use high-quality linear supplies of SELV type and/or EN 60601-1 compliant.
Note: If your power supply does not meet these requirements, the detector
performance specifications cannot be guaranteed.
Trigger Connector
The trigger circuitry inside the detector is shown in the figure below.