Up-, downloading and deleting
The File Access Control dialog box enables up-, downloading and deleting of defect
map files.
Upload
To upload a customer defect map file from the host to the detector, the
Factory/Custom
Upload
selector in the
File Access Control
menu has to be set to
Customer
before
opening the
File Access Control
menu through
Upload/Download File
.
Select Type Miscellaneous in the first drop list and in the second drop list the Defect
Map with appropriate Resolution/Full Well mode. Press the
Browse…
button to select
a defect map file on the host and press
Upload (to Camera)
to start the transfer to the
detector. The upload will be confirmed with the message:
File Upload Completed.
When attempting to upload a new defect pixel map, the detector will check the file
header. In case the header is incorrect, uploading the file will be blocked with error
message:
The selected file write access to the device failed!