36
User Manual, Rev 02
Rad-icon 2329 GigE Detector
Teledyne DALSA
Row
#
Time
Reset
Integration
Time
Frame
(
N
)
Frame
(
N
+
1
)
Trigger Input
Trigger Output
Image array readout
A
C
Pixel Array
1
2
Readout
D
B
3
4
E
Timing diagram in Snapshot Extended mode
Offset images can be acquired by putting the detector in to free running mode with
the same Extended Exposure settings.
Summing process
In FullFOV 1x1 and FullFOV 2x1 mode the summing scheme is maximal 4.
This means after 4 frames are summed one output image is streamed out. Original
data is 14 bits from the detector. After summing the result will be 16 bit.
In FullFOV 2x2 the summing scheme is maximum 2.
This means after 2 frames are summed one output image is streamed out. Original
data is 15 bits (in binning mode) from the detector. After summing the result will be
16 bit.
Frame rates
Full FOV 1x1
Full FOV 2x2
Full FOV 1x2
A
Trigger input active
B
Detector read out
125ms
62.5ms
62.5ms
C
X-ray active
Extended Exposure
setting
Extended
Exposure setting
Extended Exposure
setting
D
Dummy read (reset)
= B
= B
= B
E
Minimum period time
500ms
125ms
250ms