12
User Manual, Rev 02
Rad-icon 2329 GigE Detector
Teledyne DALSA
!
Ethernet Connector
Ethernet Connection LED
Steady orange indicates that an Ethernet connection is successfully established.
Data Transmission LED
Flashing green indicates that the detector is transmitting or receiving data.
Detector Status LED
The detector is equipped with an orange/green LED used to display the status of the
detector’s operation. The table below summarizes the operating states of the detector
and the corresponding LED states.
Color Of Status LED
Meaning
1
Orange
Power applied, Standby mode
2
Green
Detector is operational
3
Off
No power
Power Connector
Description
1
Supply voltage (+12- 28V / 20W)
2
Ground
Shielding (chassis)
The detector requires a single voltage input (+12-28V / 20W). There is a protection for
misconnection, via a resettable fuse. When the fuse has been activated (isolating),
power cycling will return it to conductive state again thus restoring normal operation
(assuming power has been properly applied).
WARNING: When setting up the detector power supplies follow these guidelines:
Apply the appropriate voltages.
The power supply must be able to deliver higher peak current of 3A during start up.
1
2
Pin 1 indicator
Rear View
Plug
Power Connector