Piranha4 8k Color Camera
User’s Manual
79
Teledyne DALSA
03-032-20274-00
The illumination, lens magnification, and focus should be set up as per you
application.
Getting the magnification right is best accomplished by setting the object-to-sensor
distance. Use the formula
lens focal length x (2 + 1/magnifi magnification)
to calculate this distance. Magnification equals the sensor pixel size (7.5 µm) / (your
object pixel size in um).
The approximate location of the sensor position is at the first groove in the side of
the case, back from the front face.
4. Camera Timing & Control
It is easiest and quickest to evaluate the camera using the internal timing setups for line
rate and exposure time.
Since we recommend starting with Camera Link medium mode, set a suitable line
rate less than 20 KHz, using the ‘ssf’ command.
If this line rate is too slow for your application, you will get a compressed image in
the scan direction. This should not be a problem for a basic evaluation.
You can set the exposure time using the ‘set’ command. Ensure that the exposure
time period is not greater than the period of the line rate minus 2.5 µsec.
The camera will indicate an error if you select an exposure time that is too long. The
minimum exposure time is 2.6 µsec.
Set your camera direction using the ‘scd’ command. Refer to the Camera direction
diagram in this manual for a definition of ‘forward’ and ‘reverse’.
5. Acquiring an Image
You can now begin imaging. Unless you have an application employing lots of light, the
image is likely to be too dark.
Use the preamp gain to increase the image dynamic (x2 or x4)
Use the system gain to adjust the camera output to achieve the desired response.
The system gain range is from 1x to 3.99x.
Once you have a suitable response, you can now focus the lens.
If you are using white LEDs, your image will have reasonable color reproduction if
you enter the provided WhiteLED color matrix.
The image may be darker at the edges due to lens vignetting, but this will be
improved once the camera is calibrated.
Calibration is performed using a white reference where your object is normally
located.
o
Use a white material that has no texture, such as a non-glossy plastic.