- Do not fold or bend the hose.
- For upper arm or thighs put the cuffs on the thickest areas. Turning the
inflated cuffs or changing the application areas with inflated cuffs may
damage the skin tissues on the application areas.
- Use the minimum pressure to stop bleeding to minimize the damage of
blood vessel, nerve, muscle or skin.
- If you can not deflation even pressing the Deflation button after surgery , remove
the hose from the device and forcibly deflated.
- The duration to stop bleeding shall not exceed generally 60 minutes for
upper body and 90 minutes for lower body. If more time is required due to
inevitable reasons, let blood flow for 15 minutes to minimize the inflammatory
reason of skin tissues.
- Keep the cuffs after washing and suitable sterilizing after application.
- Keep monitoring body temperature, blood pressure and pain of a patient
during hemostasis.
- When using two double cuffs, connect one cuff with the red hose and grey
hose to ports of 1CH(red) first and then another double cuff with blue hose
and grey hose to 2CH(blue). If Change button is pressed, the cuff bladder
connected with gray hose is activated.
- For using LOP sensor, trim the nails or toenails of a patient. If a patient has
a nail polish, remove it because it can interrupt LOP sensor to recognize.
- A cuff shall be the exclusive cuff for the model of DAESUNG MAREF. Do not use
any cuffs from other manufacturers or for other models of DAESUNG MAREF.
- Do not operate this device before putting on a cuff. Putting the inflated cuffs
may damage the cuffs and strain a body. Furthermore, putting the inflated
cuffs may decrease the efficiency of hemostasis because the cuffs may not
stick to the application area of a patient sufficiently.
- Do not press the deflate button during surgery.
(There is a risk of bleeding when the cuff is deflated.)