7-2-4. Button check
7-2-5. Sensor, Solenoid, Leak check
Display
Status
Test method
Condition that
button is not
pressed firmly.
- Verify if selected button indicates
on Display or not and make sure
if it is matched with images on
the left.
- PRESSURE, INFLATE, TIME,
DEFLATE, CHANGE, LOP,
RECORD, POWER, ALARM
button can be verified.
Condition that
button is pressed
firmly.
Display
Status
Check completion with good condition
Check failure Error condition
Condition that Check is not proceeded
* Power goes out after power button is pressed for 3 seconds.
- Connect 4 Cuffs into CH 1, CH2 before testing.
- Press the Alarm + Power button altogether then Sensor, Solenoid, Leak check mode
will show.
- Test will be proceeded 1 to 4CH accordingly.
- Press the PRESSURE UP BUTTON (red) button in order to check next chamber that
is being tested. Standard check order starts from 1CH to 4CH accordingly.
35
UserManual_DTS-3000_170223_인증용.indd 35
2017-02-23 오후 2:06:08