Product use and procedure
Use this device in accordance with the instruction from a doctor.
Select proper application area, types of cuffs and pressure level in consideration of a patient’s
status and operation area. Improper application may have adverse impact to patients during
operation.
1) Connect the parts as described from Step No. 1 to 5 above in Application and
Installation section.
2) Put the LOP sensors to the index finger or the second toe of the limb where the cuff is
applied.
3) Touch the LOP button long (3 seconds), and LOP starts measuring and displays the
proper pressure value for a patient.
4) For using the recommended pressure value by LOP sensor, touch the Inflate button to
apply the pressure to the cuff. Or set the pressure value as desired using the Pressure
Control button without using the recommended pressure value.
5) After the pressurization is complete , remove the LOP sensor worn by the patient.
5-1. Before using a device
5-2. Use and install
5-3. LOP Application
5
1) Connect the cuff and the air hose.
2) Put a bandange on the application area of a patient and put the cuff on it.
3 Connect the power cord to the device.
4) Push the ON/OFF switch at the back of the device to ON. Touch the Power button on
the front.
5) Set the pressure and time in accordance with a patient’s status and operation area.
6) Touch the Inflate button long (3 seconds) to apply the pressure to the cuff.
7) When “OK” message is displayed on the LCD screen, the pressure is completely
applied to the cuff. At this point, check the blood flow at the end of limbs and start
operation.
8) For changing the hemostasis area while using the double cuff, touch the Cuff Change
button long (3 seconds) to change the pressure area.
9) For deflating the cuff after the completion of operation, touch the Deflate button long
(3 seconds) to deflate the cuff.
10) After the air is completely deflated from the cuff, disconnect the power cable and touch
the power button to shut off the power completely.
* Even when power is forcibly off, the cuff is not deflated.
(When power cord is pulled off, the device works with the battery so that power is
not turned off).
* Do not press the deflate button during surgery.
(There is a risk of bleeding when the cuff is deflated.)
27
UserManual_DTS-3000_170223_인증용.indd 27
2017-02-23 오후 2:06:07