6
1CH LOP
Button
Then, LOP automatically estimates and notifies the proper
pressure for a patient. To start, press the Inflate button to
activate. The icon below is activated on the LCD
during LOP measurement. LOP icon color is changed as
shown below.
7
2CH
Pressure
Control
Button
Control of Pressure of 2CH
The previous setting is saved and displayed. The pressure
can be adjusted from 20 to 700mmHg depending on patient’s
status and application areas.
Set the pressure using the Pressure button ( , ).
Press for about 3 seconds. The pressure can be adjusted
during operation. (Unit : 5mmHg (Long push), 1mmHg (Short
push))
Furthermore, SET is displayed on the pressure value before
pressure is applied. SET is changed to RUN when pressure is
supplied and the current pressure value is displayed.
8
2CH
Inflation
Button
Inflation of the cuff of 2CH
Press the Inflate button long (3 seconds) to apply preset
pressure to a cuff.
When pressure is applied to a cuff, LCD displays .
When pressure reaches to the preset value, is
displayed.
9
2CH Time
Setting
Button
Time setting of 2CH
The previous value is saved and displayed. It can be adjusted
to 1 to 240 minutes depending on patient’s status or operation
types. Press the Time Setting button ( , ) to set the time.
User can change the set pressure during operation. (Unit :
5min (Long push), 1min (Short push)). The time displayed
during operation indicates the elapse time. Time can be
adjusted during operation using the button ( , ).